CCX 70
V1.1 | 29 April 2016
Water current to o lo w
If the water current is too low, electrochemical attacks at the anode can occur.
The easiest way to reduce the water resistance is to add a small amount of tap water or min-
eral water to the cooling water. Alternatively, a small quantity of copper salt (sulfate, nitrate)
can provide the necessary conductivity.
Water current to o high
The power delivered by the source is given by the anode current and anode voltage. However,
the power supply is limited by the amount of emission current it can deliver. If too much current
is being lost as water current, the power output from the source drops.
In order to reduce the water current, you should add deionized or distilled water. If you are
using tap water and the water current is too high, you need to change the water supply—a
closed circuit system offers the best solution in this case.
High water current can also be caused by contaminants in the cooling water. In this case, per-
form some cleaning cycles with tap water to wash the cooler and the water lines, then refill the
cooler with deionized or distilled water.
14
Содержание CCX 70
Страница 1: ...CCX 70 Cooling Water Controller User Manual V1 1 29 April 2016...
Страница 4: ...CCX 70 V1 1 29 April 2016 7 4 Interior Details 32 7 5 Specifications 33 Index 35 iv...
Страница 8: ...CCX 70 V1 1 29 April 2016 This page intentionally left blank 4...
Страница 13: ...9 CCX 70 V1 1 29 April 2016 Figure 1 Connection scheme for water cooling...
Страница 14: ...CCX 70 V1 1 29 April 2016 This page intentionally left blank 10...
Страница 22: ...CCX 70 V1 1 29 April 2016 This page intentionally left blank 18...
Страница 26: ...CCX 70 V1 1 29 April 2016 This page intentionally left blank 22...
Страница 32: ...CCX 70 V1 1 29 April 2016 This page intentionally left blank 28...
Страница 38: ...CCX 70 V1 1 29 April 2016 This page intentionally left blank 34...
Страница 40: ...This page intentionally left blank CCX 70 V1 1 29 April 2016 36...