Sample Extractor Assembly Operation
Note: Make sure that the front plate is closed and the inert gas/vacuum break valve is closed.
Activate the auxiliary vacuum pump to decrease the pressure in the
1.
antechamber.
After the gauge stabilizes at the maximum position, allow the vacuum pump to
2.
operate for an additional 15 minutes.
Open the antechamber isolation door. With locking handle in the up, or
3.
unlocked position, the antechamber door can be slid open.
When antechamber isolation door is fully open, lower the handle to lock the
4.
isolation door in position.
Manipulate the selector arm to select and extract the vial(s) of choice. Vials
5.
are securely held in spring loaded jaws. Push the button to open the jaws and
release to close them.
Place the sample(s) in the antechamber directly beneath the stoppering ram.
6.
Shut the isolation door then lower the handle to seal it closed.
7.
Raise handle up to
unlock the door
Push to
open the jaws
Remove shroud
to stopper vials