The OMD-625 is equipped with two programmable alarm relay contacts rated at 5A @ 230V AC,
labelled as ALARM 1 and ALARM 2 in the menu. The two alarms set points are user adjustable and
can be set either as HI or LOW, and can also be disabled or selected with a delay mode through the
menu.
To set the Alarm as ON/OFF, HI or LOW, DISABLE or DELAY select the MENU button on the overlay,
select which alarm you wish to configure and scroll down to the feature you want to change. Hit the
ENTER button and use the UP / DOWN arrows to make your next selection until the alarm is config
-
ured how you want it.
To set the alarm value; enter the menu and the alarm you wish to configure (ALARM 1 or ALARM 2.
Use the UP / DOWN arrows until your set point is displayed. If the display is showing Percent values
(%) use the DOWN button until the drop into the PPM Values.
To wire the alarms, open the explosion proof enclo
-
sure (declassifying the area if necessary) and look
at pin J2.
3.7
Alarm Configuration
Operation
OMD-625 Oxygen Analyzer
Designed, Tested and Assembled in California, USA