67
CALCulate:SCALe:FUNCtion?
Parameters:
Selected operation: {DB|DBM|PCT|SCALe}
Example:
Enable the DB calibration function for the next measurement reference:
CALC: SCAL: FUNC DB
CALC: SCAL: STAT ON
7.2.14.4 CALCulate:SCALe:GAIN
Description:
Sets the gain value M of the calibration function in CALCulate:SCALe:FUNCtion.
Syntax:
CALCulate:SCALe:GAIN {<gain>}
CALCulate: SCALe: GAIN?
Parameters:
gain
Example:
CALC: SCAL: GAIN 100
7.2.14.5 CALCulate:SCALe:OFFSet
Description:
Sets the offset value B of the calibration function in CALCulate:SCALe:FUNCtion.
Syntax:
CALCulate:SCALe:OFFSet {<offset>}
CALCulate:SCALe:OFFSet?
Parameters:
offset
Example:
CALC: SCAL: OFFSet 5
7.2.14.6 CALCulate:SCALe:REFerence
Description:
Sets the reference value for the PCT function in CALCulate:SCALe:FUNCtion.
Syntax:
CALCulate:SCALe:REFerence {<reference>}
CALCulate: SCALe: REFerence?
Parameters:
reference
Example:
CALC: SCAL: REF 100
7.2.14.7 CALCulate:SCALe:REFerence:AUTO
Description:
Enable or disable the automatic reference selection for the dB calibration function.
Syntax:
CALCulate:SCALe:REFerence:AUTO {ON|1|OFF|0}
CALCulate: SCALe: REFerence: AUTO?
Parameters:
{ON|1|OFF|0}
Example:
CALC: SCAL: REF: AUTO ON
7.2.14.8 CALCulate:SCALe[:STATe]
Description:
Enable or disable the calibration function
Syntax:
CALCulate:SCALe[:STATe] {ON|1|OFF|0}