7.2.4.1 [SENSe:]VOLTage:AC:BANDwidth
Description:
Sets the AC filter cutoff frequency for AC voltage measurements.
The instrument uses three different AC filters that allow you to optimize low frequency accuracy or reduce AC
settling time after changing the input signal amplitude. Based on the cutoff frequency specified by this command,
the instrument selects a slow (3 Hz), medium (20 Hz) or fast (200 Hz) filter. Specify the lowest frequency you
want to encounter. Lower frequencies result in longer settling times.
Syntax:
[SENSe:]VOLTage:AC:BANDwidth {<filter>|MIN|MAX}
[SENSe:]VOLTage:AC:BANDwidth? [{MIN|MAX|DEF}]
Parameters:
<filter>:{3|20|200}. Default: 20
Example:
VOLT:AC:BAND 200
7.2.4.2 [SENSe:]VOLTage:AC:SPEEd
Description:
Sets the measurement speed for AC voltage measurements.
The measurement speed is limited by the filter cutoff frequency. The 3Hz filter only supports slow (SLOW), the
20Hz filter supports slow (SLOW) and medium speed (MEDium), and the 200Hz filter supports slow (SLOW),
medium speed (MEDium) and fast (FAST).
Syntax:
[SENSe:]VOLTage:AC:SPEEd {<speed>|MIN|MAX}
[SENSe:]VOLTage:AC:SPEEd? [{MIN|MAX|DEF}]
Parameters:
<speed>:{SLOW|MEDium|FAST}. Default: MEDium
Example:
VOLT:AC:SPEE SLOW
7.2.4.3 [SENSe:]VOLTage:AC:SECondary
Description:
Select the auxiliary measurement function for the AC voltage measurement.
Please use DATA2? to read auxiliary measurement results.
Syntax:
[SENSe:]VOLTage:AC:SECondary {OFF|BEForemath|FREQuency}
[SENSe:]VOLTage:AC:SECondary?
Parameters:
{OFF|BEForemath|FREQuency}
Example:
VOLT:AC:SEC FREQ
7.2.4.4 [SENSe:]VOLTage:{AC|DC}:RANGe
Description:
Select a fixed range for AC and DC voltage measurements
Syntax:
[SENSe:]VOLTage:{AC|DC}:RANGe {<range>|MIN|MAX}
[SENSe:]VOLTage:{AC|DC}:RANGe? [{MIN|MAX|DEF}]
Parameters:
AC:<range>:{100m|1|10|100|750}. Default: 10
DC: <range>:{100m|1|10|100|1000}. Default: 1000
Example:
VOLT:AC:RANG 10