25/08/2004 ©
Söring
Service-Manual Arco- and RF-units
1
not valid for A-3000/ -2000
2
not valid for Arco-2000
-83-
3
not valid for MBC600
4
not valid for Arco-1000
Power diagram Bipolar Cutting
2,3,4
0
20
40
60
80
100
120
140
25
50
100
200
500
Resistance in Ohm
Power in Watt
10W
25W
50W
100W
130W
Power diagram Bipolar Coagulation
2,4
0
20
40
60
80
100
120
25
50
100
200
500
Resistance in Ohm
Power in Watt
10W
25W
50W
100W