1.
Evaluation of test functions at start up
1.1
External watchdog defective
1.2 Internal
watchdog
defective
1.3
CRC check of EPROM showed errors
1.4
Patter-test of RAM failed; RAM defective
2.
Read-out of measuring channels without signal application at idle
2.1
Limits: min. 7 max. 12
Idle meas. Channel I80 (current of high power supply) outside of limits
2.2
Limits: min. 0 max. 11
Idle meas. Channel PWR_L (RF-output power) outside of limits
2.3
Limits: min. 0 max. 11
Idle meas. Channel B_REL_TST (test channel bipolar relay) outside of limits
2.4
Limits: min. 0 max. 12
Idle meas. Channel OSC_VAL (test channel oscillator) outside of limits
2.5
Limits: min. 7 max. 12
Idle meas. Channel B_M_VAL (Bipolar-monopolar-recognition) outside of limits
2.6
Limits: min. 0 max. 12
Idle meas. Channel NE_VAL (Meas. Channel PE-recognition) outside of limits
3.
Application of reference voltage on selected measurement channels
3.1
Limits: min. 173 max. 190
Measurement reference voltage I80, outside of limits
3.2
Limits: min. 147 max. 172
Measurement reference voltage PWR_L (RF-output power Low I
2
C-ADC)
3.3
Limits: min. 147 max. 172
Measurement reference voltage B_M_VAL
4.
Determination of offset values in measuring channels with high precision
4.1
Limits: min. 0 max. 5
Offset determination I80
4.2
Limits: min. 0 max. 5
Offset determination U80 (Output voltage high power supply)
4.3
Limits: min. 1 max. 16
Offset determination PWR (RF-output power CPU-ADC)
4.4
Limits: min. 1 max. 83
Offset determination PWR_L (RF-output power Low I
2
C-ADC)
4.5
Limits: min. 6 max. 57
Offset determination PWR_H (RF-output power High I
2
C-ADC)
4.6
Limits: min. 0 max. 9
Offset determination HF_VOLT (RF-output voltage)
5.
Check of measuring oscillator
5.1
Limits: min. 166 max. 198 OSC_VAL outside of limits
6.
Test power supply watchdog
6.1
Limits: min. 0 max. 1
Trigger power supply watchdog, output voltage 0 Volt
6.2
Limits: min. 54 max. 82
Trigger power supply watchdog, low output voltage
6.3
Limits: min. 0 max. 1 power supply watchdog defective
Do not trigger power supply watchdog, same voltage selection as 6.2, measured:0V
7.
Test of selected output voltages of high power supply
7.1
Limits: min. 0 max. 1
Trigger power supply watchdog, 0 Volt
7.2
Limits: min. 0 max. 4
Bit 0, power supply voltage outside of limits
7.3
Limits: min. 3 max. 7
Bit 1, power supply voltage outside of limits; power supply defective