Performance Verification
WFM 1125 Option 0A/0B/0C/0D Service Manual
4-15
12. Observe the Bowtie display on the oscilloscope and confirm the skew is
within 3 ns (the marker on the display indicates 1 ns).
13. Change the oscilloscope zoom setting to off.
14. On the TSG1001 front panel press the COLOR BARS button to set the
COLOR BAR to 100 %.
15. Change the BNC cable from ANALOG OUT P
b
connector to ANALOG
OUT P
r
connector.
16. Change the oscilloscope settings to:
Waveform:
MATH1 to off and CH2 to on
. . . . . . . . . . . . . . .
Vertical scale:
10 mV/div
. . . . . . . . . . . . .
Vertical offset:
–350 mV
. . . . . . . . . . . .
17. Repeat steps from 5 to 12.
This test confirms the amplitude, rise time and fall time of the AUDIO output.
This test can be skipped in case of Option 0B or 296M format mode of Option
0C.
The following equipment is required.
Equipment Required
HDTV Signal Generator (TSG1001 and PSC1125)
Oscilloscope
Digital Audio Monitor
Two parallel signal cables
Three 75
W
BNC cables
BNC T connector
Two 75
W
terminators
Connections.
Refer to Figure 4–4 for connections..
Using the parallel signal cables supplied with the PSC1125, connect between the
G/Y on the TSG1001 and Y on the PSC1125, and between the B/Pb on the
TSG1001 and P
B
/P
R
on the PSC1125.
Using the 75
W
BNC cable, connect the OUT1 connector on the PSC1125 rear
panel to the CH A connector on the waveform monitor rear panel.
Connect AUDIO CHA connector on the waveform monitor rear panel to the
oscilloscope CH1 input using the 75
W
BNC cable and the BNC T connector.
Audio output
Содержание Tektronix WFM 1125
Страница 4: ......
Страница 6: ......
Страница 14: ...Service Safety Summary viii WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 20: ...Introduction xiv WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 21: ...Specifications ...
Страница 22: ......
Страница 39: ...Operating Information ...
Страница 40: ......
Страница 42: ...Preparation for Use 2 2 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 56: ...Using the Menu 2 16 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 61: ...Theory of Operation ...
Страница 62: ......
Страница 68: ...Theory of Operation 3 6 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 69: ...Performance Verification ...
Страница 70: ......
Страница 96: ...Performance Verification 4 26 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 97: ...Adjustment Procedure ...
Страница 98: ......
Страница 112: ...Adjustment Procedures 5 14 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 113: ...Maintenance ...
Страница 114: ......
Страница 122: ...Removal and Installation Procedures 6 8 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 145: ...Options ...
Страница 146: ......
Страница 149: ...Electrical Parts List ...
Страница 150: ......
Страница 152: ...Electrical Parts List 8 2 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 153: ...Diagrams ...
Страница 154: ......
Страница 159: ...Mechanical Parts List ...
Страница 160: ......
Страница 168: ...10 8 WFM 1125 Option 0A 0B 0C 0D Service Manual Please cut away this area below the lines ...
Страница 171: ......
Страница 172: ......