Maintenance
6-4
WFM 1125 Option 0A/0B/0C/0D Service Manual
Inspect and clean the waveform monitor as often as operating conditions require.
The collection of dirt on components inside can cause them to overheat and
breakdown. (Dirt acts as an insulating blanket, preventing efficient heat
dissipation.) Dirt also provides an electrical conduction path
that
could cause an
waveform monitor failure, especially under high-humidity conditions.
CAUTION.
Avoid the use of chemical cleaning agents which might damage the
plastics used in this waveform monitor. Use only deionized water when cleaning
the menu buttons or front-panel buttons. Use a 75% isopropyl alcohol solution as
a cleaner and rinse with deionized water. Before using any other type of cleaner,
consult your Tektronix Service Center or representative.
Inspection Ċ Exterior.
Inspect the outside of the waveform monitor for damage,
wear, and missing parts, using Table 6–2 as a guide. The waveform monitor that
appears to have been dropped or otherwise abused should be checked thoroughly
to verify correct operation and performance. Immediately repair defects that
could cause personal injury or lead to further damage to the instrument.
Table 6-2: External inspection check list
Item
Inspect for
Repair action
Cabinet, front panel, and
cover
Cracks, scratches, deformations,
damaged hardware or gaskets.
Repair or replace defective
module.
FrontĆpanel knobs
Missing, damaged, or loose
knobs.
Repair or replace missing or
defective knobs.
Connectors
Broken shells, cracked insulation,
and deformed contacts. Dirt in
connectors.
Repair or replace defective
modules. Clear or wash out dirt.
Carrying handle, bail,
cabinet feet.
Correct operation.
Repair or replace defective
module.
Accessories
Missing items or parts of items,
bent pins, broken or frayed
cables, and damaged connectors.
Repair or replace damaged or
missing items, frayed cables, and
defective modules.
Cleaning Procedure Ċ Exterior.
To clean the instrument exterior, do the following
steps:
1. Remove loose dust on the outside of the waveform monitor with a lint free
cloth.
2. Remove remaining dirt with a lint free cloth dampened in a general purpose
detergent-and-water solution. Do not use abrasive cleaners.
Inspection and Cleaning
Procedures
Содержание Tektronix WFM 1125
Страница 4: ......
Страница 6: ......
Страница 14: ...Service Safety Summary viii WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 20: ...Introduction xiv WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 21: ...Specifications ...
Страница 22: ......
Страница 39: ...Operating Information ...
Страница 40: ......
Страница 42: ...Preparation for Use 2 2 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 56: ...Using the Menu 2 16 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 61: ...Theory of Operation ...
Страница 62: ......
Страница 68: ...Theory of Operation 3 6 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 69: ...Performance Verification ...
Страница 70: ......
Страница 96: ...Performance Verification 4 26 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 97: ...Adjustment Procedure ...
Страница 98: ......
Страница 112: ...Adjustment Procedures 5 14 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 113: ...Maintenance ...
Страница 114: ......
Страница 122: ...Removal and Installation Procedures 6 8 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 145: ...Options ...
Страница 146: ......
Страница 149: ...Electrical Parts List ...
Страница 150: ......
Страница 152: ...Electrical Parts List 8 2 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 153: ...Diagrams ...
Страница 154: ......
Страница 159: ...Mechanical Parts List ...
Страница 160: ......
Страница 168: ...10 8 WFM 1125 Option 0A 0B 0C 0D Service Manual Please cut away this area below the lines ...
Страница 171: ......
Страница 172: ......