Adjustment Procedures
WFM 1125 Option 0A/0B/0C/0D Service Manual
5-9
Vertical offset:
–350 mV
. . . . . . . . . . . .
21. Align the –350 mV level of the signal with the center graticule on the
oscilloscope.
22. Change the oscilloscope CH2 vertical offset to 350 mV.
23. Verify and adjust the A10 R542 to align the 350 mV level with the center
graticule on the oscilloscope.
24. Change the oscilloscope CH1 vertical offset to –350 mV.
25. If the –350 mV level is within
"
0.5 division from the center graticule, go to
next step. If not, go back to step 21.
26. Press the TIMING SIGNALS button on the TSG1001 front panel to set the
TIMING SIGNALS to BOWTIE.
27. Change the oscilloscope settings to:
Waveform:
CH1 and CH2
. . . . . . . . . . . . . . .
Vertical scale:
100 mV/div for CH1
. . . . . . . . . . . . .
100 mV/div for CH2
Vertical offset:
0 V for CH1 and CH2
. . . . . . . . . . . .
Horizontal scale:
1
m
s/div
. . . . . . . . . . .
Record length :
5000 points
. . . . . . . . . . . .
28. Press the MATH REF button on the oscilloscope front panel, and then press
the main and side menu buttons in the following order to create a Math
waveform:
Change Math waveform definition
³
Dual Wfm Math
³
Set 1st Source to
[Ch1]
³
Set operator to [–]
³
Set 2nd Source to [Ch2]
29. Change the oscilloscope settings to:
Waveform:
CH1 and CH2 to Off
. . . . . . . . . . . . . . .
Zoom:
On
. . . . . . . . . . . . . . . . . . .
Vertical scale:
100 mV/div (5.0 X Vert) for Math1
. . . . . . . . . . . . .
Horizontal Scale:
5
m
s/div (0.2X Horz)
. . . . . . . . . .
30. Observe the Bowtie display on the oscilloscope and verify and adjust the
A10 DL210 so that the skew is within 1 ns (the marker on the display
indicates 1 ns).
Содержание Tektronix WFM 1125
Страница 4: ......
Страница 6: ......
Страница 14: ...Service Safety Summary viii WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 20: ...Introduction xiv WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 21: ...Specifications ...
Страница 22: ......
Страница 39: ...Operating Information ...
Страница 40: ......
Страница 42: ...Preparation for Use 2 2 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 56: ...Using the Menu 2 16 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 61: ...Theory of Operation ...
Страница 62: ......
Страница 68: ...Theory of Operation 3 6 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 69: ...Performance Verification ...
Страница 70: ......
Страница 96: ...Performance Verification 4 26 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 97: ...Adjustment Procedure ...
Страница 98: ......
Страница 112: ...Adjustment Procedures 5 14 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 113: ...Maintenance ...
Страница 114: ......
Страница 122: ...Removal and Installation Procedures 6 8 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 145: ...Options ...
Страница 146: ......
Страница 149: ...Electrical Parts List ...
Страница 150: ......
Страница 152: ...Electrical Parts List 8 2 WFM 1125 Option 0A 0B 0C 0D Service Manual ...
Страница 153: ...Diagrams ...
Страница 154: ......
Страница 159: ...Mechanical Parts List ...
Страница 160: ......
Страница 168: ...10 8 WFM 1125 Option 0A 0B 0C 0D Service Manual Please cut away this area below the lines ...
Страница 171: ......
Страница 172: ......