1-3 (E)
MAV-555
Microsoft(R) Windows 95
(C)Copyright Microsoft Corp 1981-1995.
C:\>
ping
\
192.168.0.1
"
C:\>
ping
\
192.168.0.1
"
Pinging 192.168.0.1 with 32 bytes of data:
Reply from 192.168.0.1: bytes=32 time=2ms TTL=254
Reply from 192.168.0.1: bytes=32 time=1ms TTL=254
Reply from 192.168.0.1: bytes=32 time<10ms TTL=254
Reply from 192.168.0.1: bytes=32 time=1ms TTL=254
C:\>
C:\>
ping
\
192.168.0.1
"
Pinging 192.168.0.1 with 32 bytes of data:
Request timed out
Request timed out
Request timed out
Request timed out
C:\>
Connection (excluding PU-109 board)
Connect the Ethernet terminal of the rear panel to a personal computer using an Ethernet cross cable.
Preparation
1.
Set the IP address and the sub-net mask of the MAV-555 to the factory settings.
m
.
The IP address and sub-net mask can be confirmed in set-up menu No. 220.
.
The factory settings are as follows:
IP address :
192.168.0.1
Sub-net mask : 255.255.255.0
2.
Set the IP address and the sub-net mask of the personal computer as follows:
IP address :
192.168.0.2
Sub-net mask : 255.255.255.0
3.
Start up the personal computer.
4.
Wait for the MS-DOS prompt (or command prompt) to appear.
5.
Input [ping 192.168.0.1].
6.
Confirm that the reply is returned normally.
Message example (normally)
Message example (abnormally)
7.
If the reply is not returned normally, input [arp -d 192.168.0.1] and perform from step 5 again.
(There can be a case that the APR table information is left remained when two or more MAV-555s
are connected continuously.)
1-2. Acquiring the Logs