4-8 (E)
DWT-B30
Measurement procedure
1. Set this unit to the desired frequency. (For the setting of frequency, refer to the Operating Instructions
of this unit.)
2. Turn on the power of this unit to enter the test mode. (Refer to Section 1-5.)
3. Select SIGNAL SELECT and set to RAWDATA PN9.
4. Set the frequency of DWR-R03D to the same frequency as this unit.
5. Locate DWR-R03D and this unit so that the distance between them is approximately 1 m.
6. Measure
BER.
Specification: BER = 1E
_
5 or less
Setting of measuring device (in the case of KBM6010)
Output frequency:
128 kHz
Data
pattern:
PN9
Measurement
range:
1E
_
6
Clock/data/synchronized
mode:
10111
4. Check of spectrum mask (UC/CE model only)
Required equipment
.
Spectrum analyzer
Setting of measuring device
.
SPAN:
2 MHz
.
RBW, VBW: 1 kHz
.
LEVEL:
3 dBm (setting: 2 mW)
10 dBm (setting: 10 mW)
14 dBm (setting: 25 mW)
.
TRACE:
Peak HOLD
Содержание DWT-B30
Страница 4: ......
Страница 32: ......
Страница 48: ......
Страница 51: ...5 3 E DWT B30 4 Click the DOWNLOAD button 5 Click the Run button 6 Click Download and install this feature ...
Страница 65: ...5 17 E DWT B30 5 When the installation is completed click the Close button The installation has been completed ...
Страница 72: ...5 24 E DWT B30 5 Restart this unit 6 Click Troubleshoot 7 Click Advanced options 8 Click Startup Settings ...
Страница 95: ...i DWT B30 Revision History Date History Contents 2019 12 1st Edition 9 932 731 01 ...
Страница 96: ......
Страница 98: ...Sony Corporation DWT B30 CE DWT B30 CN DWT B30 J DWT B30 UC J E 9 932 731 01 Printed in Japan 2019 12 08 2019 ...