5.8.9.2. API-Evaluation: Sizing Reflector
Maximize signal from reflector
being evaluated setting it’s
amplitude between 70 to 90% of
the
A-Scan
height then enter into
Freeze Peak
mode as it is
described in the paragraph 5.8.6.1
of the operating manual and click
on
– the
Draw Gate
mode (
Draw A
) is active so it is
possible to manipulate
Gate A
over
the entire
A-Scan
area
Mark the signal peak envelop then
click on
The size
D
of reflector being
evaluated according to API 5UE
practice will be determined and
indicated upon