6.2.7.2. Filling Data Logger with Measurements Results
Click on
to start capturing of the thickness reading within entire series of measurements into a
database file
1 through 4 characters determine the
address of cell for storing
single point
measurement
(thickness reading
D
accompanied with
A-Scan
); each
character is accompanied with the
numerical value. For example for
2D
data logger format will consist of 2
characters
X
and
Y
, each with the
number. To select the character, for
which the numerical address should
be modified click on it (this highlights
the selected character) then change
the numerical value through clicking
on either
or
. It should be
noted that in order to accelerate the
access to the selected cell every digit
composing numerical value may be
accessed directly through clicking on it
– the selected digit is highlighted)
To store
single point measurement
into the cell click on
Free cell is represented by the following
save / open
controls:
Cell already filled with
single point measurement
is represented by the following
save / open
controls:
To preview
single point measurement
stored in the cell click on
To overwrite
single point measurement
stored in the cell with a new one click on
6.2.7.3. Exporting Data Logger File from ISONIC utPod to Computer
Connect
ISONIC utPod
to computer
and run
ISONIC utPod for PC
SW
(refer to chapter 8 of the operating
manual). In the
ISONIC utPod
start
screen appearing on the computer’s
screen click on
In the next screen appeared click on