ISONIC utPod from Sonotron NDT - Operating Manual – Revision 1.18 - Page 61 of 94
6.2.2.2. Automatic Calibration
Place probe onto the sample of material with known thickness allowing receiving of several back echoes,
then provide gating (
Gate A
) according to paragraph 6.2.1 of the operating manual, then click on
,
the screen as below appears:
Key in the Thickness of the material, on which
calibration is performed
The corresponding
USVelocity
value is defined and
displayed automatically
Click on
to use automatically found
USVelocity
value as default, as a result the precise material
thickness result is displayed: