Chapter 5:
Cleaning and Disinfecting
49
Chapter 5: Cleaning and Disinfecting
Universal Precautions
SonoSite recommends that personnel who have regular exposure to medical devices returned for service practice
“universal precautions.” Universal precautions are an approach to infection control. Those servicing this product
should follow the prescribed standards for their area.
Receipt of Suspected Contaminated Materials
SonoSite recommends that personnel who have regular exposure to medical devices returned for service practice
“universal precautions.” Universal precautions are an approach to infection control. Those servicing this product
should follow the prescribed standards for their area.
If visual inspection suggests possible contamination when opening a product returned for service, take proper
steps to contain the contamination. Wear necessary Personal Protective Equipment (PPE) (gloves, masks, and
gowns) when opening or examining a suspect package.
Before transfer to a service area, label the suspect package “contaminated” and seal it to prevent exposure.
Discard any packing materials removed from a package suspected of contamination in a biohazard container.
Discard any contaminated materials received with the product in an appropriate biohazard container.
Contaminated materials may include biohazardous waste and sharps.
Maintain a disinfecting agent in case any work surface is contaminated. The recommended agent is 0.5% sodium
hypochlorite (bleach) solution. To prepare the agent, mix one part household bleach (5.25% - 6% sodium
hypochlorite) to nine parts water. Spray or wipe the solution onto the work surface and allow to air dry.
Please use these recommendations when cleaning or disinfecting your ultrasound system, transducers, and
accessories. This chapter assists in effective cleaning and disinfection, but it is also intended to protect the system
and transducers against damage during cleaning or disinfection.
For more information about cleaning or disinfecting solutions or ultrasound gels for the transducer, call SonoSite
technical support or your local representative. For information about a specific product, call the product
manufacturer.
Recommended Disinfectants
For a list of disinfectants recommended for use on the system and transducers, see the
MicroMaxx Ultrasound
System User Guide
.
Содержание MicroMaxx
Страница 1: ...MicroMaxx Ultrasound System Service Manual ...
Страница 6: ...vi ...
Страница 10: ...4 Chapter 1 Introduction ...
Страница 56: ...50 Chapter 5 Cleaning and Disinfecting ...
Страница 60: ...54 Chapter 6 Troubleshooting Troubleshooting Flow Diagrams Display Figure 6 2 Display Flow Diagram ...
Страница 61: ...Chapter 6 Troubleshooting 55 Control Panel Figure 6 3 Control Panel Flow Diagram ...
Страница 62: ...56 Chapter 6 Troubleshooting System Figure 6 4 System Flow Diagram ...
Страница 63: ...Chapter 6 Troubleshooting 57 Battery Figure 6 5 Battery Flow Diagram ...
Страница 72: ...66 Chapter 7 Replacement Procedures Figure 7 10 Power Supply Screws Screws 7 ...
Страница 85: ...Appendix A Parts List 79 Figure A 1 Power Supply P05471 Figure A 2 Speaker Assembly P03872 ...
Страница 89: ...Appendix A Parts List 83 Figure A 8 Main PCB Assembly P05472 ...
Страница 92: ...86 Appendix B Service Event Report ...
Страница 98: ...92 Index ...