Содержание MICROPLOTTER Proto

Страница 1: ...MICROPLOTTER PROTO MANUAL REVISION 1 3 SONOPLOT INC 3030 LAURA LANE SUITE 120 MIDDLETON WI 53562 608 824 9311 CONTACT SONOPLOT COM...

Страница 2: ......

Страница 3: ...le Application 19 One component oligonucleotide microarray 19 Wires from nanometallic inks 27 SonoGuide 29 Overview 29 Main interface 29 Directional controls 31 Toolbar items 32 Saving positions 33 Ro...

Страница 4: ...Dispenser refurbishing 58 Revision History 60 3...

Страница 5: ...Microplotter among others are trademarks of SonoPlot Inc in the United States and other countries Other parties trademarks or service marks are the property of their respective owners and should be tr...

Страница 6: ......

Страница 7: ...liters of fl uid to a surface in order to create features as small as 20 micrometers microns wide due to its use of a novel form of fl uid ejection based on ultrasonic pumping The dispenser at the cor...

Страница 8: ...pot on a surface but not deactivating the dispenser after fl uid contact has been made with the surface Instead the dispenser is moved along the surface in a straight line leaving behind a smooth line...

Страница 9: ...be of much use without a precision positioning system to place that dispenser at select locations and do so in an automated fashion The robotic positioner of the Microplotter system occupies most of i...

Страница 10: ......

Страница 11: ...d and optics system already mounted iMac control computer with keyboard and mouse Control electronics A box containing cables for attachment of the various components Additionally SonoGuide the primar...

Страница 12: ...Rear view of control electronics box Power cable Positioning stage X Y and axis cables These are interchangeable and can be attached in any order USB cable to control computer 9...

Страница 13: ...e electronics box into a USB port on your computer and also insert the USB cable from the CCD camera to a different USB port on your computer This step is shown below on a Mac computer The process is...

Страница 14: ...er connectors to the control electronics Installing the SonoGuide software The SonoGuide software is available on both the MacOS platform and Windows platform Follow the steps below to install the Son...

Страница 15: ...older SonoGuide has now been installed to your MacOS computer Windows Go to www sonoplot com downloads and click on the SonoGuide for Windows link Download the SonoGuideInstaller exe Once downloaded r...

Страница 16: ...Select the components to install 13...

Страница 17: ...Select where you would like to create SonoGuide s shortcut SonoGuide is ready to be installed 14...

Страница 18: ...Click Install to start installing SonoGuide Once the installer has fi nished installing SonoGuide click on Finish SonoGuide has been installed and is ready to use 15...

Страница 19: ...ting the exposed glass needle to any surface as even a slight touch can shatter the fragile tip To load a dispenser cartridge carefully place the notched side of the dispenser onto the printhead as sh...

Страница 20: ...ront of the control box When the system is powered on the green power light and orange COMM light will be illuminated on the front of the box Double click on the SonoGuide icon to open the controlling...

Страница 21: ...Fully installed Microplotter Proto system 18...

Страница 22: ......

Страница 23: ...Microplotter instrument to lay down continuous lines These examples assume that the system has been properly installed as described in Chapter 2 One component oligonucleotide microarray To begin turn...

Страница 24: ...port microscope slides either replace the deck with one that does or use a different substrate material After loading the slide clip into place a new dispenser if one is not already in the system Conn...

Страница 25: ...opening the Camera Controls from the windows options and adjusting the sliders in the dialog that will appear Close this window when fi nished Now that the dispenser and camera are set up properly it...

Страница 26: ...ver shooting the surface too far will damage the glass needle Change the large step size to 1000 microns one millimeter and more fi nely position the dispenser relative to the surface of the microscop...

Страница 27: ...ve been entered press the Calibrate surface cant button to begin the surface topography measurement The dispenser will move down until it contacts the surface the position will be measured and then th...

Страница 28: ...be easy to return to later To do this click on the button in the Positions window This will create a new position and leave you able to edit its name Type in Above slide and hit Return when you are d...

Страница 29: ...to add a new solution to the list Enter in a name of Solution 1 and hit Return when you are done Click Set Custom Position to set this up as a custom position using the current coordinates When the g...

Страница 30: ...ve the dispenser 2 4 millimeters to your right and towards you to better center the pattern on the slide With the addition of fl uid to the dispenser there is the possibility that its vibrational prop...

Страница 31: ...ositioner to its safe resting location and remove the fi nished microarray from the deck of the Microplotter system Wires from nanometallic inks A typical microelectronic application of a Microplotter...

Страница 32: ...SonoGuide window to switch back to that application Manually move the dispenser over to the well containing the ink to be deposited and down into the ink until the tip is below the fi ll level of the...

Страница 33: ...windows and arrange these windows in any fashion the user fi nds useful Main interface Upon starting SonoGuide it will check with the robotic positioner of the Microplotter system to determine its st...

Страница 34: ...been homed it will prompt you that it is about to do so Warning Clear the working area of the Microplotter system and make sure that no one is near the volume covered by the robotic arm as the positio...

Страница 35: ...starting with the directional controls Directional controls The window for the directional controls appears as follows These controls allow you to move the positioner manually in a stepwise fashion or...

Страница 36: ...eir home points Warning The home move travels over a large distance and can be very fast so make sure that all objects and personnel are clear of the positioner s working range before clicking on this...

Страница 37: ...on highlight a position by clicking on it and press Crosshairs This will cause the dispenser to pick up to a set height the setting of which will be explained in the next section move over to the corr...

Страница 38: ...is being moved manually and when the movement is controlled by the computer during the automatic deposition of a pattern For example it might be desirable to have faster speeds and accelerations set...

Страница 39: ...By enabling this option the system will cause the dispenser to approach the surface in smaller steps and use a more sensitive detection routine The downside to this is that the approach to the surfac...

Страница 40: ...loaded in this manner Once a pattern has been loaded it will be previewed in the middle of the main interface A crosshair indicates the current position of the dispenser relative to the rest of the pa...

Страница 41: ...eir name Only the enabled and visible layers will be printed These layers can be associated with solution wells that have been laid out in SonoGuide by clicking next to them in the Solution Well colum...

Страница 42: ...stay at the last location in a pattern when done as opposed to the normal process of returning to the start on print completion with the second option The third option lets the system optimize the pat...

Страница 43: ...me display from the CCD camera pointed at the tip of the fl uid dispenser The high magni fi cation optics surrounding the CCD have a fi eld of view that is approximately 1200 microns 1 2 mm across and...

Страница 44: ...save the snapshot to Additionally you can select the image format of the snapshot from a range of different types with many quality options to choose from At the top of the main user interface there a...

Страница 45: ...be used with Exposure to tune the camera input for different lighting conditions Brightness The overall brightness of the camera image too high and it can reduce the contrast of the image too low and...

Страница 46: ...praying action within a hollow cavity Settings for controlling the behavior of the dispenser can be found under Dispenser Show dispenser controls or the Control dispenser toolbar item In the default v...

Страница 47: ...rom this view you can manually activate the dispensing or spraying process by clicking on the Dispense or Spray buttons respectively The dispensing or spraying action will take place for the duration...

Страница 48: ...d to be recalibrated due to changes in its contents or drift in the system The connection of a dispenser to the system is not automatically detected Open the Calibrate Dispenser toolbar item to begin...

Страница 49: ...on dialog either choose the menu item Dispenser Perform Dispenser Calibration or the Calibrate dispenser toolbar item Solution well locations The Microplotter has the ability to dispense multiple solu...

Страница 50: ...tion has been selected its properties will appear on the Solution Settings window In addition to solution wells locations for rinse solvents and waste wells can be set These can be shared between solu...

Страница 51: ...it takes to spray all of that solution out of the dispenser and the number of rinse cycles required to fully clean out that solution from the dispenser Each solution can be set to either return solut...

Страница 52: ...ime of a dispenser can be prolonged by preventing positioning overshoot that might damage a dispenser s fragile tip The application of this surface sensing has been described in the Main interface and...

Страница 53: ...tical directions After the sensing run has completed the cant of the surface in X and Y will be calculated stored and reported in this window This is expressed as a slope the number of microns to rise...

Страница 54: ...ration will be used for every pattern printed after this point until you clear the calibration To clear the calibration click on the Clear rotational calibration button at the bottom of the dialog box...

Страница 55: ...face sensor as described in the previous section When retracted from the surface the signal should sharply drop off once again If this behavior is not observed repeat the calibration and try again Thi...

Страница 56: ...ngle next to the Advanced label This will expose a series of realtime readouts from the ultrasonic frequencies used to determine when contact has been made with a surface This can be used by SonoPlot...

Страница 57: ...uts Find Surface Ctrl Cmd F Set A1 position Ctrl Cmd 1 Move to desired position Ctrl Option click Go Home Fn Left arrow MacOS Go Home Home key Windows When using Trackball Hold SHIFT to move 10X When...

Страница 58: ......

Страница 59: ...de of the dispenser s tip Other polymers may harden to the point where solvent is unable to dissolve the polymer permanently sealing it shut SonoGuide gives the error message Positioner is not connect...

Страница 60: ...e on the dispensing tip Sonoplot suggests moving the tip down to a short distance above the surface then stepping down in 10 micron steps until a fl uid bridge is made between the tip and the surface...

Страница 61: ...the sensor diagnostics dialog by choosing the Dispenser Show Diagnostics menu item as described in Chapter4 This window displays the realtime readout from the surface sensor A red line indicates the...

Страница 62: ......

Страница 63: ...as 10 Viscosity 450 cP Positioning 31 x 31 x 7 cm X Y Z axes 10 m resolution Calibration Automatic surface height calibration Camera USB 3 0 with digital video capture Computer Included running Mac OS...

Страница 64: ......

Страница 65: ...pipette from the wide end by scribing the glass and breaking it Use a scoring wafer such as the Sutter ceramic scoring wafer www sutter com part number CTS Reassemble the dispenser by replacing the RJ...

Страница 66: ...59...

Страница 67: ...chapter to include new version UI features Version 1 2 Added the polygon pattern element to Chapter 9 Version 1 1 Revised the SonoGuide chapter to include the new pattern preview window and DXF import...