Technical data
The abbreviation „i.v.” in the basic uncertainty
definition means the indicated value.
Constant voltage measurement (on the dis-
play):
* - from 6,0V for Sr. no. up to 681793
Additionally the voltage is indicated by the diode
bar for the following values: 12, 24, 50, 120, 230,
400, 690V along with the signalling of the voltage
polarization (the diode „+” or „-” is lit). The diode
indicator functions also without batteries.
Input resistance
Uin
Rin
12V, 24V, 50V
~ 6k
Ω
120V
~ 20k
Ω
230V
~ 70k
Ω
400V
~ 150k
Ω
690V
~ 240k
Ω
Alternating voltage measurement within the
range between 20 and 400Hz (on the display):
* - from 6,0V for Sr. no. up to 681793
Additionally the voltage is indicated by the diode
bar for the following values: 12, 24, 50, 120, 230,
400, 690V along with the signalling of alternating
voltage (diodes „+” and „-” are lit simultaneously).
The diode indicator functions also without batter-
ies.
The measurement voltage frequency for the diode
bar: 15…400Hz.
Resistance measurement:
Range
Resolution
Basic
uncertainty
0...1999
Ω
1
Ω
±(3% i.v. + 8
digits)
Other data:
a) Kind of insulation: double, in accordance with
EN 61010-1
Range
Resolution
Basic
uncertainty
+1,5/-2,5*
...49,9V
0,1V
50...750V
1V
±(2% i.v. + 3
digits)
Range
Resolution
Basic
uncertainty
1,5*...49,9V
0,1V
±(3% i.v. + 4 dig-
its)
50...750V
1V
±(2% i.v. + 3 dig-
its)