65
SBP
SOMMER sensor that supports the SBP-protocol (via RS-485)
The sensor address is set in
, and the position of the measurement value
within the output string is assigned in
MIO
SOMMER sensor that supports the MIO-protocol (via RS-485)
The sensor address is set in
, and the position of the measurement value
within the output string is assigned in
SYS
System variable
key A
Number of times the MRL-7 is activated via its keyboard.
key S
Number of times the MRL-7 is activated via its keyboard with the
special key sequence
+
.
manual
If the selected value has been changed manually, the updated value is
stored.
exep. A
Diagnostic variable
exep. B
Diagnostic variable
light
Ambient light intensity (0…3000 Lux)
+Sup V
Supply voltage
+Bat V
Voltage of internal battery
5V sup
Voltage of internal 5V bus supply
mo CSQ
Modem signal strength (not available for MRL-7O)
mo excep.
Modem exception message (not available for MRL-7O)
mo srv1
Modem data server 1 (not available for MRL-7O)
mo srv2
Modem data server 2 (not available for MRL-7O)
mo srv3
Modem data server 3 (not available for MRL-7O)
Switch
Status of switched current output
Message
Number of successful message transmissions since the last storage
event. The message number defined in
entered in
RECYC
“Recycled” variable
The number of a referenced variable is set in
Example:
Variable nr.
03
is set to analogue input 1 (
AIN1
) and represents the raw value.
Variable nr.
12
is set to RECYC and its
03
to process the raw value. If set-
0.5
, variable nr.
12
returns half of the raw measurement value.
RECYCM
multiple “Recycled” variables
The numbers of referenced variables is set in
GPS
Output of the integrated GPS-receiver (only available in MRL-7GPS)
Long
The longitude of the data loggers position
Lat
The latitude of the data loggers position
Altitude
The altitude of the data loggers position
Sat
The number of satellites in view of the GPS-receiver
Attention:
must be on for use of GPS-functionality!
S-NUM
The number of the selected SDI-12 or RS-485 sensor, e.g.
3
. Valid values for the SDI-12 address are
0…9, a…z and A…Z. The RS-485 device address may take a value 1…98.