Levelogger Field Notes

Before Deployment

Before deployment, make sure you do the following:
•  Program your Levelogger, using Levelogger Software, with the correct 

project identification, memory mode, sampling regime, time, etc.

Note:

 It is useful to synchronize the times of all Leveloggers and  

 

Barologgers being used for the same project. 

•  Set a future start time, or start the Levelogger if deploying on a wireline/

Kevlar rope (Leveloggers can be started after deployment if using a Direct 

Read Cable)

•  Determine borehole depth to ensure Levelogger does not touch the bottom 

of the well (avoid submergence in sediment)

•  Record the deployment depth of your Levelogger
•  Use a Solinst Model 101 or 102 Water Level Meter to take a manual depth 

to water measurement that will be used to verify Levelogger readings 

•  Determine the minimum and maximum expected water levels, as 

Leveloggers must remain submerged for the entire monitoring period, and 

Barologgers must not be submerged

Deployment

•  Deploy your Levelogger and Barologger using a direct read cable for down-

well communication, or use an inexpensive wireline or Kevlar cord.

Note:

 For information on other types of installations, see the latest  

 

Levelogger User Guide.

•  Install the Barologger in a similar thermal environment as the Levelogger
•  The Barologger should be suspended beyond the frost line and deep enough 

to avoid large temperature fluctuations

•  Ensure the Barologger location is vented to atmosphere

Tip

It is recommended to 

take a manual water 

level measurement 

before installing a 

Levelogger, shortly after 

installation, periodically 

during your monitoring 

interval, and at the end 

of your measurement 

period.  Use these 

measurement to verify 

Levelogger readings, and 

for data adjustments 

later on. Ensure you 

take manual readings 

as close in time as 

possible to a scheduled 

Levelogger readings.

After Deployment

After deployment, make 

sure you do the following:
•  Take a manual depth to 

water measurement after 

the well has stabilized 

(approximately 10 

minutes)

•  Take another manual 

depth to water 

measurement just before 

removing the Levelogger 

from the well

 

Wireline/Cord  

Deployment

 

Direct Read  

Deployment

 

Well Caps

The Model 3001 2" 

(or 4” with Adaptor) 

Well Cap Assembly 

provides a secure 

method of installing 

your Levelogger using 

wireline/Kevlar cord or 

Direct Read Cables.

- 6 -

Содержание Levelogger Series

Страница 1: ...High Quality Groundwater and Surface Water Monitoring Instrumentation Quick Start Guide Levelogger Series...

Страница 2: ...ardware Wizard 2 The Found New Hardware Wizard will give the option to connect to Windows Update to search for software components select No not at this time 3 Two options are provided i If you have t...

Страница 3: ...ttings icon will create a template Datalogger Settings Window Programming the Levelogger Note Clicking on the embedded in the software will provide you with a short explanation of that feature e g Com...

Страница 4: ...urements to adjust your data to depth to water readings Note The default directory for downloaded and saved data is in the Data folder C Program Files Solinst Levelogger 4_0 Data Data is saved as a xl...

Страница 5: ...ater lying above the Levelogger pressure sensor AND the barometric atmospheric pressure acting on the water surface To compensate for barometric pressure fluctuations and get true height of water colu...

Страница 6: ...cation or use an inexpensive wireline or Kevlar cord Note For information on other types of installations see the latest Levelogger User Guide Install the Barologger in a similar thermal environment a...

Страница 7: ...Levelogger from freezing To avoid icing freezing and transducer damage the easiest method is to lower the transducer to a point in the water column below the frost line or ice formation depth In water...

Страница 8: ...e device is installed on 3 Reason Another device shares the same Com Port or is causing a communication conflict Solution Ensure that software for PDA or other devices which automatically synchronize...

Отзывы: