Troubleshooting

Levelogger PC Software:

1. You must have administrator privileges to install software on a computer.
2. Windows 7, 8, and 10 Operating Systems support Levelogger Software.

AquaVent data has been accidentally erased:

If the AquaVent has been restarted and the old data has not been saved, select the download option  

“Recover Previous Log”.  This will download your previous data set.

Try the following if your AquaVent is not communicating with Levelogger PC 
Software, or you are getting a communication error message:

1. Make sure you have installed the batteries correctly: 

•  Ensure proper polarity, paying attention to the positive {+} and negative {-} symbols etched in 

the holder 

•  Ensure proper alignment; the spring contacts in the housing should be lined up with the 

protruding contacts on the bottom of the battery holder

2. Check the Vented Cable connections to the AquaVent logger and Wellhead. Ensure they are 

lined-up correctly and there is no dirt or debris in the connection.

3. Make sure you are using the correct Wellhead Connector Cable to connect to the PC. 
•  You must be using a USB Connector Cable connected to the 10-pin Solinst Protocol connection 

on the Wellhead. The cable and connection are colour-coded orange. 

4. Check that the USB Connector Cable is connected to the same Com Port that is 

 

chosen in the upper middle of the Main Window of the Levelogger Software.

5. Check the Com port settings. They should be as follows:

•  Bits per second: 9600
•  Data bits: 8
•  Parity: None
•  Stop bits: 1
•  Flow control: None (This may have been set to Xon/Xoff – change it to None, Select [OK]   

and back out of this pathway.)

6. Try communicating with another AquaVent logger, Wellhead, Vented Cable or USB Connector 

Cable. 

7. Try using a different computer.

8. If problem persists, contact Solinst.

To reset the AquaVent battery indicator, after replacing the Wellhead batteries:

Go to the Diagnostic Utility in Levelogger Software, click “Run Diagnostics”, then click “Reset 

AquaVent Battery Indicator”.

Note:

  For diagnostic purposes, you may want to install the end of the Vented Cable with the serial number 

to the surface connection at the Wellhead. 

Содержание AquaVent 3500

Страница 1: ...High Quality Groundwater and Surface Water Monitoring Instrumentation Quick Start Guide 3500 AquaVent Please read this document carefully before using the AquaVent...

Страница 2: ...Wellhead Vented Cable AquaVent Logger 10 Pin SP Connection Alignment socket in the Vented Cable connectors Alignment pin in the Wellhead and logger connectors Coupling Connector Body Line up the alig...

Страница 3: ...gger Settings Window Note Clicking on the embedded in the software will provide you with a short explanation of that feature e g Com Port Slate Mode Time Synchronization etc Starting and Stopping Aqua...

Страница 4: ...view rate is set independently of the logging period of the AquaVent and does not interfere with internal logging taking place To take a reading at any specific time click and that reading will be add...

Страница 5: ...reference datum in the Manual Data Adjustment option in the Software Data Wizard AquaVent loggers detect both water W and barometric pressure B However a vent tube to surface allows barometric effect...

Страница 6: ...g 4 using an adaptor However other installations are acceptable if proper precautions and recommendations are followed Leave at least 2 50 mm of slack in the Vented Cable at surface to allow you to li...

Страница 7: ...ing blue red USB connection has been dropped Flashing yellow The Wellhead is communicating with the PC Software Maintenance As with any monitoring project you should select the proper equipment and de...

Страница 8: ...e bottom of the battery holder 2 Check the Vented Cable connections to the AquaVent logger and Wellhead Ensure they are lined up correctly and there is no dirt or debris in the connection 3 Make sure...

Отзывы: