AquaVent Quick Start Guide

In-Field Communication

There are two SP Wellhead Cable options for communicating with AquaVent loggers in the field.

•  The USB Connector Cable allows you to connect to a field laptop for communication using Levelogger 

Software (see image on Page 1).

•  The App/DataGrabber Connector Cable allows you to connect to a Levelogger App Interface for 

communication using the Solinst Levelogger App, or to a DataGrabber for downloading data to a  

USB device. 

- 7 -

Note

The SPX Wellhead has 

additional LED definitions 

related to MODBUS and  

SDI-12 operation. 

For information on 

communication using 

MODBUS or SDI-12 Protocols, 

see separate User Guides.

App/DataGrabber  

Connector Cable

DataGrabber

LED: Wellhead Connection Status 

The AquaVent Wellheads have a multicolour LED light which 

indicates the status and operation of the AquaVent:

•  Fast multicolour cycle 3 times:  Wellhead is powering-on
•  Fast 10 cycles alternating blue/green: Wellhead has successfully    

connected to the PC using the USB Connector Cable

•  Fast 10 cycles alternating blue/red: USB connection has been  

dropped

•  Flashing yellow: The Wellhead is communicating with the PC Software

Maintenance

As with any monitoring project, you should select the proper equipment and determine a maintenance 

schedule based on the environment specific to your application. 
To ensure the vent tube remains dry, the AquaVent loggers and Wellheads contain permanent 

desiccants and hydrophobic membranes at the connections to the Vented Cable—there is no need 

to replace them. Please see our technical bulletin on 

Vented Cable Design, Use, and Maintenance 

for more details. 
Other maintenance tips include (see the AquaVent User Guide for more details):

•  Replace the o-rings at the Vented Cable connections on the AquaVent logger and the Wellhead. 
•  Keep the dust caps on all of the connections, when not in use. 
•  Store the AquaVent logger in its case, and the Vented Cable on its spool when not in use.
•  Ensure connections are clean before assembling. Clean the AquaVent logger as required. 
•  Replace the Wellhead batteries as required (four user-replaceable 1.5V AA lithium batteries). 

AquaVent Sampling Rate

Battery Life Estimate

8 readings per second

10 days

1 reading per second

3 months

1 reading per minute 

8 years

Levelogger App Interface

Содержание AquaVent 3500

Страница 1: ...High Quality Groundwater and Surface Water Monitoring Instrumentation Quick Start Guide 3500 AquaVent Please read this document carefully before using the AquaVent...

Страница 2: ...Wellhead Vented Cable AquaVent Logger 10 Pin SP Connection Alignment socket in the Vented Cable connectors Alignment pin in the Wellhead and logger connectors Coupling Connector Body Line up the alig...

Страница 3: ...gger Settings Window Note Clicking on the embedded in the software will provide you with a short explanation of that feature e g Com Port Slate Mode Time Synchronization etc Starting and Stopping Aqua...

Страница 4: ...view rate is set independently of the logging period of the AquaVent and does not interfere with internal logging taking place To take a reading at any specific time click and that reading will be add...

Страница 5: ...reference datum in the Manual Data Adjustment option in the Software Data Wizard AquaVent loggers detect both water W and barometric pressure B However a vent tube to surface allows barometric effect...

Страница 6: ...g 4 using an adaptor However other installations are acceptable if proper precautions and recommendations are followed Leave at least 2 50 mm of slack in the Vented Cable at surface to allow you to li...

Страница 7: ...ing blue red USB connection has been dropped Flashing yellow The Wellhead is communicating with the PC Software Maintenance As with any monitoring project you should select the proper equipment and de...

Страница 8: ...e bottom of the battery holder 2 Check the Vented Cable connections to the AquaVent logger and Wellhead Ensure they are lined up correctly and there is no dirt or debris in the connection 3 Make sure...

Отзывы: