
135
26. CHECKS AND ADJUSTMENTS
X
PROCEDURE Adjustment
8. Store values X2 and Y2
Press
[OK]
to set the horizontal
angle to 0°. “Take F2” is displayed.
9. Rotate the top of the instrument
through 180°.
10. Wait a few seconds for the display
to stabilize, then store the
automatically compensated
angles X1 and Y1.
Press
[YES]
to store tilt angles X1
and Y1. The new correction
constant is displayed.
11. Confirm that the values are in the
adjustment range.
If both correction constant are
within the range 400±30, select
[YES]
to renew the correction
angle. <Instr. const> is restored.
Continue to step 12.
If the values exceed the
adjustment range, select
[NO]
to
cancel the adjustment and restore
<Instr. const>. Contact your
Sokkia agent to perform the
adjustment.
30R.book 135
ページ
2003年12月16日 火曜日 午後5時15分
Содержание 30R Series
Страница 8: ...vi 30R book vi ページ 2003年12月16日 火曜日 午後5時15分 ...
Страница 172: ...29 REGULATIONS 164 CE Conformity Declaration 30R_29 FM 164 ページ 2003年12月18日 木曜日 午後5時11分 ...
Страница 173: ...165 29 REGULATIONS 30R_29 FM 165 ページ 2003年12月18日 木曜日 午後5時11分 ...
Страница 177: ...30R book 1 ページ 2003年12月16日 火曜日 午後5時15分 ...
Страница 178: ...8th ed 13 0408 Printed in Japan 2002 SOKKIA CO LTD 30R_C3C4 FM 2 ページ 2004年8月23日 月曜日 午後5時43分 ...