G r a s e b y M R 1 0 I n s t r u c t i o n M a n u a l
G r a s e b y M R 1 0
S m i t h s M e d i c a l
Страница 1: ...d and understood the contents of the complete manual including the warnings and cautions and have been trained in the correct use of the product Smiths Medical cannot be held responsible for the accuracy and any resulting incident arising from information that has been extracted from this publication and compiled into the users documentation This publication maybe subject to revision and it is the...
Страница 2: ...Instruction Manual Graseby MR10 Neonatal Respiration Monitor ...
Страница 3: ...out the prior permission of Smiths Medical Smiths Medical PM Inc N7 W22025 Johnson Drive Waukesha WI 53185 1856 U S A European Representative Smiths Medical International Limited Colonial Way Watford Hertfordshire UK WD24 4LG Australian Representative Smiths Medical Australasia Pty Ltd 61 Brandl Street Eight Mile Plains QLD 4113 Australia www smiths medical com Registered in England Company number...
Страница 4: ...ve been fully trained in its correct set up and operating procedures Failure to observe this warning could affect the functions of the Monitor and may cause infant injury or death 6 WARNING The Monitor is not suitable for use on infants with pacers Failure to observe this warning may affect the functions of the Monitor and may cause infant injury or death 7 WARNING The Monitor should only be used ...
Страница 5: ...eath 19 WARNING Excessive movement may adversely affect the Monitor s performance Care must be taken when used in an ambulatory situation and the Monitor placed in a secure position free from excessive movement suspendusingthestrapsupplied Failuretoobservethiswarningcouldaffect the functions of the Monitor and may cause infant injury or death 20 WARNING TheGrasebyRespirationSensor partnumber0108 0...
Страница 6: ...of the Monitor can cause incorrect or loss of monitoring resulting in infant injury or death 27 WARNING To avoid possible malfunction of the Monitor do not expose the MonitortoX rays gammarays ionisingradiation strongelectricormagnetic fields such as diathermy or mobile telephones Malfunction of the Monitor cancauseincorrectmonitoringorlossofmonitoringresultingininfantinjury or death 28 WARNING Fo...
Страница 7: ... or subject the Monitor to temperatures in excess of 45 C 113 F Excessive temperatures may cause damage to the Monitor 3 CAUTION Do not use unapproved cleaning agents 4 CAUTION To prevent serious damage to the Monitor it must not be immersed in any liquids or exposed to strong organic solvents Wipe off spills immediately Do not allow fluid or residues to remain on the Monitor Additionally the Moni...
Страница 8: ...or 1 6 Monitor set up 1 7 Switching on the Graseby MR10 1 7 Silencing the clicking sound 1 8 Changing the alarm delay 1 8 Checking alarm output 1 8 Monitoring the infant 1 8 Switching off the Graseby MR10 1 9 If the alarm sounds 1 9 Batteries 1 10 Battery alarm 1 10 Battery replacement 1 10 Faults causes and actions 1 12 Care and maintenance 1 13 Maintenance 1 13 Cleaning the Monitor 1 13 Respirat...
Страница 9: ...1 G ra s e b y M R 1 0 I n s t r u c t i o n M a n u a l G ra s e b y M R 1 0 S m i t h s M e d i c a l ...
Страница 10: ...every time the Monitor is used If you are unsure about any aspect of the Monitor s use contact your clinician WARNING The clinician should ensure that the performance offered by the Monitor is fit for the intended purpose Failure to observe this warning could affect the functions of the Monitor and may cause infant injury or death Overview of the Monitor The Graseby MR10 Neonatal Respiration Monit...
Страница 11: ...must be returned to an authorised technician for repair before being returned to clinical use Failure to observe this warning could affect the functions of the Monitor and may cause infant injury or death WARNING Before using the Monitor and its accessories ensure that you have been fully trained in its correct set up and operating procedures Failure to observe this warning could affect the functi...
Страница 12: ... by pressing the HOLD FOR OFF button Keep the button pressed until you hear a short beep If the Alarm system test is unsuccessful see page 1 12 Faults causes and actions for further information WARNING If the main audible and visual alarms are activated immediately attend to the infant This is of prime importance before any other actions are taken In the event of any alarms the infant must first b...
Страница 13: ...h the clothing 3 It is important to ensure that the Respiration Sensor connector is pushed firmly into the Monitor s connector socket using a slight twisting action Do not apply excessive pressure NOTE When attaching the Respiration Sensor pad ensure that the tape is long enough and does not interfere with the Respiration Sensor The tape should not be stretched too tightly over the Sensor pad Only...
Страница 14: ...age 1 9 If the alarm sounds Should the infant start to breath again before is pressed then the audible alarm will stop but the red alarm light will continue to flash for another 60 seconds WARNING The Monitor should only be used in the home environment under the direction of a clinician Failure to observe this warning may cause infant injury or death WARNING The Monitor must be operated within the...
Страница 15: ...ed as soon as possible by pressing the button WARNING Mobile communication devices e g mobile phones should not be used at a distance of less than one metre to the Graseby MR10 Monitor as this could affect the safe operation of the Monitor Even if these devices cause no obvious interference possible interference can only be ruled out if this advice is strictly observed Malfunction of the Monitor c...
Страница 16: ... being exerted on the keypad buttons causing it to be switched off or the settings changed Failure to observe this warning could result in infant injury or death If the alarm sounds WARNING If the main audible and visual alarms are activated immediately attend to the infant This is of prime importance before any other actions are taken In the event of any alarms the infant must first be made safe ...
Страница 17: ...uaranteed The operational life of the batteries depends on the frequency of the alarms However a service life of approximately 60 days can generally be assumed The batteries should be removed if the Graseby MR10 is not to be used for a prolonged period It is recommended that at least one full set of spare batteries is always available Battery replacement 1 To replace the batteries open the battery...
Страница 18: ...batteries can be discharged after being stored for a long time WARNING Use only type LR6 alkaline batteries Rechargeable batteries are unsuitable and must not be used as they could affect safe operation of the Monitor Failure to observe this warning could affect the functions of the Monitor and may cause infant or user injury WARNING When the Low Battery orange indicator is lit it is essential to ...
Страница 19: ...rtment Batterries discharged Fit new batteries No respiration signals Respiration Sensor not connected to the Monitor Make sure that the Sensor connector is inserted in the Monitor s connector socket using a twisting motion Sensor pad has become loose on the skin Remove the Sensor clean and dry the infant s skin Attach the Sensor pad with new hypo allergenic tape See page 1 6 Aattching the Respira...
Страница 20: ...ot be immersed in any liquids or exposed to strong organic solvents Wipe off spills immediately Do not allow fluid or residues to remain on the Monitor Additionally the Monitor is not designed to allow it to be sterilised Failure to observe these cautions may cause internal damage to the Monitor The Monitor should be wiped clean using only a cloth slightly moistened with a liquid disinfectant Neve...
Страница 21: ...en 10 sec orange 20 sec Battery indicator Orange battery nearly discharged Environmental conditions Temperature range Operating 10 C to 40 C Storage 10 C to 55 C Humidity range Operating 30 to 75 RH Storage 10 to 95 RH Pressure range Operating 700 hPa to 1060 hPa Storage 500 hPa to 1060 hPa EMC Law relating to electromagnetic compatibility Group 1 Class B emitted RF radiation to EN55011 1991 suita...
Страница 22: ...E Directive 2006 96 EC and implementing regulations all devices and service items within the scope of the Directive purchased new after August 13 2005 must be sent for recycling when ultimately becoming waste Devices and items must not be disposed of with general waste If purchased before that date they may also be sent for recycling if being replaced on a one for one like for like basis this vari...
Страница 23: ...1 1 6 G ra s e b y M R 1 0 I n s t r u c t i o n M a n u a l G ra s e b y M R 1 0 S m i t h s M e d i c a l ...
Страница 24: ......
Страница 25: ...esentative Smiths Medical Australasia Pty Ltd Brisbane QLD 4113 Australia www smiths medical com Smiths Medical part of the global technology business Smiths Group 2007 Smiths Medical family of companies All rights reserved Revision issue date September 2007 Part No 0514 0182 102 For further information please contact your local distributor or Smiths Medical direct on 44 0 1923 246434 ...