G r a s e b y M R 1 0 I n s t r u c t i o n M a n u a l
1 -
S m i t h s M e d i c a l
G r a s e b y M R 1 0
Switching.off.the.Graseby.MR10
To switch the Monitor off press the
HOLD FOR OFF
button (for about 2
seconds) until a short beep is heard.
WARNING:.The.Monitor.is.not.suitable.for.use.on.infants.with.pacers..
Failure.to.observe.this.warning.may.affect.the.functions.of.the.
Monitor.and.may.cause.infant.injury.or.death.
WARNING:..The.Monitor.is.not.suitable.for.use.with.adults.or.children.
over.the.age.of.18.months..Failure.to.observe.this.warning.may.cause.
injury.or.death.
WARNING:..The.Monitor.must.not.be.covered.by.bedding.or.other.
objects.so.that.audible.and.visual.alarms.are.impaired..Failure.to.
observe.this.warning.could.result.in.infant.injury.or.death..
WARNING:..To.prevent.the.Monitor.being.accidentally.switched.off,.it..
must.not.be.placed.within.reach.of.the.infant..Failure.to.observe.this.
warning.could.result.in.infant.injury.or.death.
WARNING:..The.Monitor.must.not.be.pushed.into.confined.spaces,.
possibly.resulting.in.pressure.being.exerted.on.the.keypad.buttons.
causing.it.to.be.switched.off.or.the.settings.changed..Failure.to.
observe.this.warning.could.result.in.infant.injury.or.death..
If.the.alarm.sounds
WARNING:.If.the.main.audible.and.visual.alarms.are.activated,.
immediately.attend.to.the.infant..This.is.of.prime.importance.before.
any.other.actions.are.taken..In.the.event.of.any.alarms,.the.infant.must.
first.be.made.safe.and.comfortable..Failure.to.observe.this.warning.
may.cause.infant.injury.or.death.
1. Go to the infant but do not panic.
2. If the infant has stopped breathing carry out the resuscitation
techniques that you have learned and immediately seek medical
assistance.
3. If the red alarm indicator is flashing but the audible alarm is silent,
this shows that a previous alarm condition was present. Press the
button to cancel the visual alarm indication and observe the infant for
a while. If an audible alarm sounds again, carry out the Alarm system
test see
page 1-5,
Alarm system test
.
4. If the audible alarm is repeated, press the
button again and
observe the infant.
5. If the infant is safe and breathing normally but the Monitor is not
indicating the infant’s breath, there is possibly a technical fault, see
page 1-12,
Faults, causes and actions
.