32
a
nnual
I
nsPectIon
and
t
estIng
☐
Battery Fallout Alarm Test
Equipment needed
• 1 CADD
™
medication cassette reservoir
or
1 CADD
®
administration set
• 1 CADD® key
• Timer
Procedure
1. Program the continuous rate to 1.0 mL/hr.
2. From the Tasks menu, select
View Advanced Tasks,
and then select
Air and Occlusion
Settings
.
Turn the air detector off.
3. Disconnect the AC adapter.
4. Attach, latch, and lock an empty CADD
™
medication cassette reservoir or CADD
®
administration set to the pump.
5. Start the pump and allow it to run for a minimum of 4 minutes.
6. Start a timer to measure the time the audible alarm is active. Open the battery door.
7. The pump should respond with a continuous 2-tone alarm. The alarm should sound for a
minimum of 2 minutes.
☐
Audible Alarm Check
Procedure
1. From the Tasks menu, select
Display and Sound Settings,
and then select
Alarm Volume.
2. Scroll and pause on the 3 alarm volumes (low, medium, and high). The pump should sound the
alarm as you pause on each volume.
3. Press
U
or
D
to verify all 3 audible alarm volumes.