14
P
umP
o
vervIew
Delivery rate
during priming
•
Standard volume administration set: approx. 250 mL/hr
•
High volume administration set: approx. 500 mL/hr
Alarm disabled
during priming
Air-In Line
Delivery Specifications
Common Delivery Specifications
Reservoir volume
0 to 9999
Programmable in 1 mL increments.
Displayed in 0.1 mL increments.
Given
0 to 99,999.99 in 0.01 unit increments
Res vol low trip
point
1 to 999 mL in increments of 1 mL
Res vol empty
alarm
•
Insistent and one time only
•
Non-insistent and repeating
Delayed start
0 to 96 hr in 5 min increments
Pump stopped
alarm
•
Informational
•
High priority
Air detector
•
On
•
Off
Air detector
sensitivity
Low Sensitivity:
Single bubble > 400 μL
High Sensitivity:
Single bubble > 150 μL
Alarm volume
•
High
•
Medium
•
Low
PM (preventive
maintenance)
reminder
Interval: 1 to 24 months in 1 month increments
Enable: On or off
Custom keypad
code
001 to 899 in increments of 1
Custom clinician
code
001 to 899 in increments of 1
Custom
administrator code
001 to 899 in increments of 1
Date format
•
US standard (month/day/year)
•
European standard (day/month/year)
•
International standard ISO 8601:2004 (year/month/day)
Time format
•
00:00 to 23:59 military
•
12-hour
am
/
pm
Downstream
occlusion
sensitivity
High Sensitivity:
When the high pressure alarm threshold is reached, the
downstream occlusion alarm is triggered immediately.
Low Sensitivity:
When the high pressure alarm threshold is reached, the
downstream occlusion alarm is delayed for 2 seconds. This allows for the
pressure to stabilize before a possible alarm. If the pressure stabilizes below
the high pressure alarm threshold before the 2 second delay is complete, the
alarm will not occur.