SMC
sierra monitor corporation
Sentry Instruction Manual - Version 6
SERVICE
Page: 65
7.3.6 DIAGNOSTIC REPORT
A Diagnostic report can be printed using diagnostic codes
0008 or 0010. Figure 7-4 is a typical report..
ID:A 11/20 15:16
S1= +0.376
BV= +2.021
BI= 0.323
SP= 0653
ST= +0.284
MV= +19.0
CS= +0.000
S2= +1.20
HV= +0.920
HI= +0.975
SP= 400
ST= +0.009
MV= +19.4
CS= +0.000
Figure 7-4
Typical Diagnostic Report
Table 7-2 is a description of the information contained in
a typical diagnostic report
Report
Description
ID:A
Controller Identification
11/20 15:16
Time report was generated
S1
+0.376
Data that follows is for Module 1
Sensor Signal is +0.376 VDC
BV
+2.021
Bridge Voltage 2.021 VDC
BI
+0.322
Bridge Current 323 mA
SP
0653
Sensor Power 653 mW
ST
+0.284
Sensor Type
(Where: .2 -.3 VDC: Combustible)
MV
+19.0
Module Voltage = 19 VDC
CS
+0.000
Calibration Switch
Off 0.000, On = 2.5 VDC
S2
+1.20
Data that follows is for Module
Sensor Signal is + 1.20 VDC
HV
+0.920
Heater Voltage Signal
HI
+0.975
Heater Current Signal
SP
400
Sensor Power 400 mW
ST
+0.009
Module Type
(Where: .11-.1 VDC: H2S)
Table 7-2
Diagnostic Report Description
Содержание 5000 Sentry
Страница 6: ......
Страница 20: ......
Страница 22: ......
Страница 29: ...SMC sierra monitor corporation Sentry Instruction Manual Version 6 Page 23 Figure 3 3 Sensor Module Mounting...
Страница 30: ...SMC sierra monitor corporation Sentry Instruction Manual Version 6 Page 24 Figure 3 4 Wiring Options...
Страница 31: ...SMC sierra monitor corporation Sentry Instruction Manual Version 6 Page 25 Figure 3 5 Typical Splice Box Wiring...
Страница 38: ......
Страница 58: ......
Страница 60: ......
Страница 90: ......
Страница 94: ......
Страница 96: ......
Страница 102: ......
Страница 124: ......