Thermo-con Rack Mount Type    

Series 

HECR

Alarm 

code

Alarm description

Operation

status

Main reason

WRN

Upper/Lower temp. limit alarm

Continue The temperature has become out of upper/lower limit range for the target temperature.

ERR01

System error 1

Stop

The internal cable of the Thermo-con has been broken due to abnormal vibration or dropping the product.

ERR02

System error 2

Stop

EEPROM data has been lost due to high level noise.

ERR03

Back-up data error

Stop

EEPROM data of the controller has been destroyed due to high level noise.

ERR11

DC power supply failure

Stop

The DC power supply has failed (due to fan stop or abnormal high temperature) or the 
thermo-module has been short-circuited.

ERR12

Internal temp. sensor high temp. error

Stop

The internal temperature sensor has become higher than high temp. cutoff setting.

ERR13

Internal temp. sensor low temp. error

Stop

The internal temperature sensor has become lower than low temp. cutoff setting.

ERR14

Thermostat alarm

Stop

The thermostat has been activated due to filter clog or fan/pump failure, etc.

ERR15

Abnormal output alarm

Continue

The temperature cannot be changed even at 100% output due to overload or disconnection 
of the thermo-module.

ERR16

Low flow rate alarm (Option)

Stop

The flow rate of the circulating fluid has dropped.

ERR17

Internal temp. sensor disconnection alarm

Stop

The internal temperature sensor has been disconnected or short-circuited.

ERR18

External temp. sensor disconnection alarm Continue

The external temperature sensor has been disconnected or short-circuited. (Only detected 
when in learning control or external tune control)

ERR19

Abnormal auto tuning alarm

Stop

Auto tuning has not been completed within 20 minutes.

ERR20

Low fluid level alarm

Stop

The amount of circulating fluid in the tank has dropped.

P V

<

E R R 1 1   W R N

S V

3 1 .   6    

°

C   # 1

3 0 .   0    

°

C  

M O D E   N o r m a l

Alarm

Operation Display Panel

This unit is equipped as standard with a function allowing 1 4 kinds of alarms to display on the LCD and can be read out by serial 
communication. Also, it can generate relay output for upper/lower temperature limit alarm and output cutoff alarm.

AT key
Used to start and stop auto tuning.

1st line
Indicates No. corresponding to the alarm 
which  arises  and  [WRN]  comes  to  light 
up when the upper or lower temperature 
limit warning occurs.

4th line
Indicates  control  operation  mode  during 
normal  operation  and  set  values  during 
setting mode selection.

2nd, 3rd line
Indicates  present  temperature  [PV]  and 
target  temperature  [SV]  during  normal 
operation.  When  the  alarm  arises,  the 
error is indicated instead and during set-
ting mode selection, the selected setting 
mode is indicated.

 key

Used to change set value in each setting 
mode.

SEL key
Used to change setting mode.

RET key
Used to fix set value or return to present 
temperature status indication.

Maintenance

Maintenance of this unit is performed only in the form of return to and repair at SMC's site. As a rule, SMC will not conduct on-site 
maintenance.

Alarm

Содержание HECR Series

Страница 1: ...18to 0 054 F 0 01 to 0 03 C Temperature stability 50 to 140 F 10 to 60 C Set temperature range 200 w Cooling capacity 440 w 100 V input 4 6 A Power consumption Energy saving Good space utilization Hei...

Страница 2: ...uid volume can be checked q Turn the power ON w Press the SEL key and adjust the temperature setting with the keys e Press the RET key to complete Canpreciselycontrolthetemperatureofaheatsourceorproce...

Страница 3: ...poses a Peltier device thermo module between the heat exchangers for the circulating fluid and facility water and controls the DC power supply to achieve the target outlet temperature of circulating f...

Страница 4: ...aling Electronic microscope Ultra sonic wave inspection machine Cooling of laser irradiated part Temperature control of X ray tube and X ray light sensing part Cooling of laser irradiated part Cooling...

Страница 5: ...pacity Page 8 Heating Capacity Page 8 Pump Capacity Thermo con Outlet Page 8 Dimensions Page 9 Operation Display Panel Page 10 Alarm Page 10 Maintenance Page 10 Options With Feet and No Rack Mounting...

Страница 6: ...ssary select the Peltier type Thermo con HEC series refer to the following or the refrigerated Thermo chiller HRS HRZ series 3 How much cooling capacity required Guide to Model Selection Cooling capac...

Страница 7: ...x 103 x 4 2 x 103 60 x 1000 167 W 200 W Cooling capacity Considering a safety factor of 20 167 W x 1 2 Water Density 1 x 103 kg m3 Specific heat C 4 2 x 103 J kg K Cooled substance total volume V Cool...

Страница 8: ...ty Note 2 0 018 to 0 054 0 01 to 0 03 C Pump capacity Refer to the performance charts Page 8 Tank capacity Approx 0 34 gal 1 3 L Port size Rc1 4 Wetted parts material Stainless steel EPDM Ceramics PPE...

Страница 9: ...300 200 100 0 0 Circulating fluid Water Ambient temperature 59 F 15 C Ambient temperature 95 F 35 F Heating capacity W 1000 900 800 700 600 500 400 300 200 100 0 0 Circulating fluid Ethylene glycol 2...

Страница 10: ...ng removing the product to from the rack Circulating fluid inlet Circulating fluid drain port Rc1 4 Circulating fluid outlet Rc1 4 Dimensions HECR002 A5 1 Power supply connector IEC60320 C14 or equiva...

Страница 11: ...disconnected or short circuited ERR18 External temp sensor disconnection alarm Continue The external temperature sensor has been disconnected or short circuited Only detected when in learning control...

Страница 12: ...kets and handles on the front side are removed as they are not necessary when the product is not mounted in a rack This option has rubber feet for installing the product on the floor E F F HECR002 A5...

Страница 13: ...e connector on the Thermo con side in position For single phase 200 VAC type Also applicable for the 100 VAC type but the connector for the user s equipment needs to be prepared by the user Part no Ap...

Страница 14: ...he product weight is held with the bottom surface of the product Use the handles on the front side of the product when installing removing the product to from the rack 1 The inlet for radiation air mu...

Страница 15: ...tank is made of resin PE Therefore the tank may be crushed if the pressure is negative Special attention must be paid if the flow rate of the circulating fluid is high To avoid getting negative press...

Страница 16: ...on or equipment in contact with food and beverages emergency stop circuits clutch and brake circuits in press applications safety equipment or other applications unsuitable for the standard specificat...

Отзывы: