Description of your controller’s symbols:
There are a number of technical markings on your controller. These can be explained as
follows:
The product is manufactured for Bio-Medical Research Ltd, Parkmore Business
Park West, Galway, Ireland.
The controller requires 1 x 3.6 volt (NiMH) DC battery pack. DC is indicated by the
symbol:
The Output Frequency indicates the number of pulses per second transmitted by the
controller. This is measured in hertz, which is indicated by “Hz”.
Output (RMSA) means the maximum output root mean square current for each channel.
Output (RMSV) means the maximum output root mean square voltage for each channel.
This symbol means “Attention, consult the accompanying documents”.
This symbol means type BF applied parts.
This symbol on your controller is to indicate conformity to the requirements of
the Medical Device Directive (93/42/EEC).
The CE mark applied also indicates
that this equipment complies with the requirements of the ROHS Directive
(2011/65/EU) for the Restriction of Hazardous Substances in Electrical and
Electronic Equipment.
Conforms to UL STD 60601-1.
Certified to CAN/CSA.
STD C22.2 NO 601.1
SN stands for ‘serial number’. On the label, on the back of the controller is the serial
number specific to this product. The letter preceding the number indicates the year of
manufacture, where ‘X’ denotes 2018, ‘Y’ denotes 2019 etc.
The garment’s batch number is represented on the garment packaging, by the number
corresponding with the LOT symbol.
Rated Outputs – Voltage/Currents: X:60
Parameter 500
Ω
1K
Ω
1K5
Ω
Output RMSV
3.8V
7V
7.9V
Output
RMSA
7.7mA 7m 5.3mA
Output Frequency
50Hz
50Hz
50-Hz
DC Component: Approx.
0 C
0 C
0 C
Positive
Pulse
Width: 100-125µs 100-125µs 100-125µs
Negative Pulse Width:
100-125µs
100-125µs
100-125µs
Interphase Interval:
100 µs
100 µs
100 µs
Attention:
This product can deliver currents in excess of 2mA/cm
2
if used with incorrect electrodes.
At the end of the product lifecycle, do not throw this product into the normal
household garbage, but bring it to a collection point for the recycling of electronic
equipment.
Some product materials can be re-used if you bring them to a recycling point. By
re-using some parts or raw materials from used products you make an important
contribution to the protection of the environment. Please contact your local
authorities if you need more information about collection points in your area.
Waste Electrical and Electronic Equipment can have potentially harmful effects on
the environment. Incorrect disposal can cause harmful toxins to build up in the air,
water and soil and can be harmful to human health.
Garment Materials:
Outer material: 100% Nylon; Inner Material: 34% Polyester, 66% Expanded Polythene;
Binding: 100% Polyester; Hook & Loop: 100% Nylon; Elasticated Straps: 85% Nylon, 15%
Spandex; Foam: EVA.