Error message (Err)
Reason
Remedy
Measuring error (body
position changed)
Too much interference (body
position changed, too much
speaking
etc.
while
measuring)
Stay
in
the
right
position for measuring
and do not speak or
move.
Measuring
error
(wrong air pressure)
Wrong air pressure (air
inflation does not reach
30mmHg in 20 seconds)
Check whether the
device is leaking air or
the wristband is not
properly fastened and
adjust accordingly.
Measuring
error
(inflation
time
limit
exceeded)
Inflation time limit exceeded
(pressure increased for 60
seconds
or
above
250mmHg )
If the wristband is not
fastened
properly
readjust it.
Measuring error (low
battery in the device)
Low voltage warning
Charge the device.
Other Faults
If the product does not work normally, first check whether there is enough power, whether
Bluetooth is connected and whether the wristband is properly fastened. If none of these apply see
"Reset".
Reset: if the device behaves oddly and no solution can be found, Insert a fine needle into
the small hole on the right of the power button (not the wide hole on its left) to reset the
device.
Product Specifications
Product Name:
XM-01 Automatic Electronic Blood Pressure Monitor
Common Name:
Automatic Electronic Blood Pressure Monitor
Model:
XM-01
Means of Measurement:
Oscillographic method
Range of Measurement:
Blood pressure: 0mmHg-260mmHg