Introduction

Checking Pump/Kit Contents

Use the following table to verify that you received all items associated with the Cat. No. ordered. If you are missing items, contact SKC 
at 800-752-8472 (U.S. only) or 724-941-9701.

If you ordered Cat. No.

Your Package Should Contain

220-1000TC

Pocket Pump TOUCH only*

 with Li-Ion battery and screwdriver, requires charger; 

see kits or charger below

220-1000TC-C

Pocket Pump TOUCH with Charger*

 includes pump as described above and single charger (cable and wall cube with 

U.S. plug)  

100-240 V

220-1000TC-S

Pocket Pump TOUCH Starter Kit*

 includes pump and single charger as described above, 3 feet (0.9 meter) of Tygon 

tubing, and collar clip with cable tie  

100-240 V

220-1000TC-K

Single Pump Kit*

 

includes pump and single charger as described above, and Type A single tube holder, in a soft-sided 

nylon carry case   

100-240 V

220-1000TC-K5

5-pack Pump Kit*

 includes 5 pumps as described above, 5 single chargers (cables and wall cubes with U.S. plug), and 

5 Type A single tube holders, in a Pelican case   

100-240 V

 

220-1000TC-K5D

5-pack Deluxe Pump Kit*

 includes 5 pumps as described above, 5 Type A single tube holders, 5 single charging cables 

(no wall cubes), one 5-port USB charging hub with power cable (U.S. plug), and 1 DataTrac Pro USB Bluetooth Adapter 
(software available via free download), in a Pelican case  

100-240 V

Pocket Pump TOUCH requires ¼-inch ID tubing.

*  Pocket Pump TOUCH pumps contain Li-Ion ba

tt

 eries and are subject to special shipping regulations.

†  Single kits are also available in hard-sided Pelican case. Contact SKC.

Quick Guide

Charging the battery: 

Insert USB connector on cable into USB port on supplied wall adapter. Plug into 100 to 240-V outlet. Hover 

magnetic charging connector on cable closely over contact on back of pump until engaged and LEDs on contact and cable light blue. 
Full charge is achieved in approximately 10 hours. 

See Charging the Battery Pack on page 3 for details.

Turning on: 

Press Power On/Off button on top of pump; the 

screen will light and Flow screen will display.

Using the touch screen: 

Use a 

fi

 ngertip to gently touch 

screen buttons in the indicated active zones.

Accessing Main Menu: 

Touch 

.

Navigating from Main Menu

Desired Operation

Button to Touch

Result

Access device settings

History, Clock, Units, and Screen sub-menus display. Touch 

buttons to view and choose setting options within each.

View information about the pump

Firmware version/date of pump manufacture, and pump 

serial number display.

Access 

fl

 ow rate, calibration, and sampling

Flow, Presets, and Advanced (constant pressure) sub-menus 

display. Touch buttons to view and choose mode or options 

under each.

Basic Pump Operation

Run pump: 

From Main Menu: Touch Sample  >Flow  >Run 

Note: 

Pressing the Power On/Off button on top of pump will dim and lock screen. To undim and unlock, press Power On/Off 

button again.

Pause pump: 

Touch Pause 

. Touch Run 

 to resume.

Stop pump: 

Touch Stop 

. Sample Summary will display. Touch 

 to return to  Main Menu.

Turning off: 

Press Power On/Off button on top of pump; the screen will go dark. The pump will power off automatically after 

fi

 ve minutes of no activity.

Active Touch Screen Areas

Содержание Pocket Pump TOUCH

Страница 1: ...fying Device Settings Clock Menu Date Time Units Menu Display Units Screen Menu Dim Security Operation Sample Menu Set Flow Calibrate and Sample 13 To set pump flow rate To set up a single sorbent tube train and calibrate Constant Flow Mode To set up a multiple sorbent tube train and calibrate Constant Flow Mode To set up a multiple sorbent tube train and calibrate Constant Pressure Mode To calibr...

Страница 2: ...e subject to special shipping regulations Single kits are also available in hard sided Pelican case Contact SKC Quick Guide Charging the battery Insert USB connector on cable into USB port on supplied wall adapter Plug into 100 to 240 V outlet Hover magnetic charging connector on cable closely over contact on back of pump until engaged and LEDs on contact and cable light blue Full charge is achiev...

Страница 3: ...ging contact but the charger and charging contact LEDs are not lit blue Check that the charging cable is connected to an AC adapter and active outlet PC laptop connected to an active power source or a portable power pack with a full charge that is compatible with charging Pocket Pump TOUCH If only the charger LED lights blue It means that there is no connection with the pump charging contact Check...

Страница 4: ...laced in an area to be known nonhazardous Afin de prévenir l inflammation d atmosphères dangereuses ne changer les batteries que dans des emplacements désignés non dangereux Maximum charge input voltage Um 5 5 VDC The battery used in this device may present a risk of fire or explosion when heated above 212 F 100 C or incinerated Replace battery with SKC Battery Pack Cat No P76303 only Use of anoth...

Страница 5: ...dicated below Navigating Menus and Screens Pocket Pump TOUCH operates through a series of menus and screens When the pump is powered on the Flow screen displays which allows you fast access to calibration and sampling For more details start at To set pump flow rate Immediately below the display are four virtual navigation buttons that access previous screen Main Menu and increase decrease values B...

Страница 6: ...d to pump Advanced Menu Set Pressure Set Duration timer Run button Menus and screens can contain the following navigational touch buttons Button General Function Check mark Saves a selected item Left and right movement Allows horizontal movement on a scale see right or moves left or right through fields activating each for entry of a value Calibration Flow Adjustment Allows selection of an adjustm...

Страница 7: ... on the back of the pump These indicate pump status at a quick glance LED Activity Description Pump Status Corner and inlet LEDs green flashing Pump running Corner and inlet LEDs red flashing Pump is in flow fault Inlet LED yellow orange flashing corner LEDs green blinking Pump in constant pressure mode working to achieve pressure setting Inlet LED will change to solid green when set pressure is a...

Страница 8: ...ce Touch Clock Touch time displayed Hour digits will flash Touch up down arrow buttons to increment decrement hours Touch right arrow to advance to minutes will flash and up down arrow buttons to adjust minutes Touch right arrow to advance toAM PM will flash and up down arrow buttons to toggleAM PM Touch check mark to accept new time and return to Clock Menu New time setting will display To change...

Страница 9: ...ll flash and up down arrow buttons to increment decrement day Touch right arrow to advance to year will flash and up down arrow buttons to increment decrement year Touch check mark to accept new date and return to Clock Menu New date will display To change date display format From Main Menu Touch Device Touch Clock Touch Date Display Format Touch the desired date display format to select and retur...

Страница 10: ...hange atmospheric pressure display units From Main Menu Touch Device Touch Units Touch ATM mbar inHg and mmHg buttons display Touch desired button to select and return to Units Menu Note Changing the display units affects only the display of atmospheric pressure on the pump screen and in Sample Summary and History Back pressure inlet pressure will always display in inH2O on the pump screen and in S...

Страница 11: ... screen backlight will stay on during the entire sample run To set Auto Lock to on or off From Main Menu Touch Device Touch Screen Touch Auto Lock Touch desired button to select and return to Screen Menu If Auto Lock is set to On the screen will lock become inactive as soon as the pump starts running a sample A lock icon and Press top button to unlock message will appear on the screen If Dim is se...

Страница 12: ...combination Note Entered digits will display as You will be prompted to confirm the passcode Touch the screen keypad to enter the same four digit number combination Upon entering the last digit the passcode will be saved and you will be returned to the Screen Menu If Secure Lock is set to On the user can lock the screen of a running pump by pressing the Power On Off button If Auto Lock is set to On...

Страница 13: ...rbent tube and insert arrow printed on tube pointing toward the pump into the rubber sleeve of a 222 Series non adjustable Low Flow Tube Holder Figure 3 2 Connect tubing from the 222 Series non adjustable low flow tube holder to the pump inlet Figure 3 3 Use a length of flexible tubing to connect the exposed open end of the representative sorbent tube in the holder to a calibrator outlet Figure 3 ...

Страница 14: ...rough all tubes to be used during the multiple tube sample run to ensure adequate airflow Note The combined flow through all tubes cannot exceed 500 ml min 2 Press tubing from the outlet unlabeled side of a Constant Pressure Controller CPC onto the pump inlet Figure 4 3 Press tubing from the 224 Series Multiple tube Low Flow Holder onto the CPC inlet labeled To Sample Figure 4 4 Break tips off a re...

Страница 15: ...s on holder operation It is good practice to run the pump for 5 minutes before calibrating the flow rate Calibrate verify pump flow rate before and after each sampling operation using the tube holder and pump to be used for sampling 1 Determine back pressure to be set in Constant Pressure Mode a Connect tubing from holder outlet to pump inlet Figure 5 b Break tips off a representative sorbent tube ...

Страница 16: ...to turn flow adjust screw on port clockwise to decrease counterclockwise to increase until calibrator displays the method specified flow rate Note The flow rate displayed on the calibrator is the flow rate through the sorbent tube 6 Repeat Steps 3 through 5 for all remaining open sorbent tubes 7 Disconnect representative sorbent tube from the calibrator Replace representative sorbent tubes with ne...

Страница 17: ... Flow Touch calibration icon Pump will run Allow pump to run for 5 minutes before performing calibration Touch up down arrow buttons to increment decrement calibration adjustment A calibration adjustment value will display as shown below left Note The flow rate displayed on the calibrator will change as a result of this adjustment The flow rate displayed on the pump will not change Touch check mar...

Страница 18: ...ag sampling a Use PTFE tubing to connect pump exhaust Figure 1 to the bag fitting See bag operating instructions b Unscrew cap on pump exhaust to remove it and screw in supplied exhaust fitting or optional Quick connect Bag Sampling Adapter accessory c Attach tubing from the bag to the fitting or adapter on the pump exhaust 3 On the pump choose from a manual sample timed sample or sample preset pr...

Страница 19: ...o Lock is set to Off the screen remains active The screen may be locked made inactive at any time during sampling by pressing the Power On Off button on top of the pump To unlock and reactivate the screen press the Power On Off button on top of the pump This feature helps to reduce tap errors during sample runs Secure Lock set to On locks inactivates the pump touch screen when the Power On Off but...

Страница 20: ...bing after sampling by using fingertips to press center ring on adapter while pulling tubing away from the adapter To set up and run a timed sample 1 Set sample duration as follows From Main Menu Touch Sample Touch Flow or Advanced Hour digit 1 will flash Touch up down arrow buttons to increment decrement hour Touch right arrow to advance to hour digit 2 will flash and up down arrow buttons to adj...

Страница 21: ...ch Presets Touch desired Preset programmed in DataTrac Pro Software and uploaded to pump Presets are created by the user in DataTrac Pro for Pocket Pump TOUCH Software and uploaded to the pump see DataTrac Pro for Pocket Pump TOUCH User Manual for details on creating presets The start date for a preset with delayed start can be changed on the pump by touching the date on the Preset screen The pump...

Страница 22: ...kinked tubing can restrict airflow and cause back pressure to build to a point at which the pump can no longer compensate flow within 5 If this condition is sustained for 3 to 10 seconds the pump will go into flow fault mode as follows 1 Pump stops running and status LEDs on pump flash red Elapsed time stops 2 An exclamation point icon appears on the display next to the flow rate 3 After 20 second...

Страница 23: ...o 877 94 and DataTrac Pro for Pocket Pump TOUCH Software which is available as a download requires Internet connection Access DataTrac Pro for Pocket Pump TOUCH Software using the following procedure 1 Check that PC meets DataTrac Pro system requirements Operating system Microsoft Windows 7 8 8 1 or 10 Required software DataTrac Installer included with installation Minimum display resolution 1024 ...

Страница 24: ...ttery pack housing with pump case b Use the back of the battery pack housing to push up the belt clip until the battery pack is aligned with the pump case and presses easily into place Ensure the battery pack housing fits flush with the pump case c Use a 1 5 mm hex key to tighten two screws on back of the battery pack housing Tighten screws in an alternating fashion d Charge the battery pack compl...

Страница 25: ... 224 26 03 224 26 04 Tube Covers A 6 mm OD x 70 mm L B 8 mm OD x 110 mm L C 10 mm OD x 150 mm L D 10 mm OD x 220 mm L 224 29A 224 29B 224 29C 224 29D Replacement Battery Pack Li Ion P76303 Single Charging Cable USB without wall cube P75739 Low Flow chek mate Calibrator 20 to 500 ml min includes a 9 volt alkaline battery and NIST traceable calibration certificate 375 00205N Replacing baƩeries with ...

Страница 26: ... below sea level Flow Fault After 3 to 10 seconds of restricted flow pump stops running elapsed time stops status LEDs flash red and pump displays fault icon After 20 seconds in fault auto restart is attempted up to 5 times unless full airflow is restored prior to that If full airflow is not corrected after 5 auto restarts within 5 minutes the pump ends the run and displays Sample Summary Display ...