![SKC Pocket Pump TOUCH Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/skc/pocket-pump-touch/pocket-pump-touch_operating-instructions-manual_1286855013.webp)
Form 40017 Rev 1906
Page 13 of 26
Operation - Set Flow, Calibrate, and Sample
Sample Menu
To set pump
fl
ow rate
From Main Menu:
Touch Sample
> Touch Flow
> Touch
fl
ow display
Touch
left/right arrow bu
tt
ons
to set gross
fl
ow se
tt
ing. Touch
up/down arrow bu
tt
ons
to
fi
ne tune
se
tt
ing.
Note:
A sustained touch on the up/down arrow bu
tt
ons will speed increment/
decrement
fl
ow se
tt
ing.
Touch check mark to accept selection and return to Flow Menu with new
fl
ow se
tt
ing displayed.
To set up a single sorbent tube train and calibrate – Constant Flow Mode
* Charge pump battery completely before calibration and sampling.
* Requires 222 Series Low Flow Holder accessory.
* Calibrate/verify pump
fl
ow rate before and after each sampling operation
using the tube holder and pump to be used for sampling.
1. Break tips o
ff
a representative sorbent tube and insert (arrow printed on
tube pointing toward the pump) into the rubber sleeve of a 222 Series
(non-adjustable) Low Flow Tube Holder
(Figure 3)
.
2.
Connect tubing from the 222 Series (non-adjustable) low
fl
ow tube holder
to the pump inlet
(Figure 3)
.
3. Use a length of
fl
exible tubing to connect the exposed open end of the
representative sorbent tube in the holder to a calibrator outlet
(Figure 3)
.
4. Set
fl
ow rate on pump
(see To set pump
fl
ow rate above)
.
5. Calibrate
fl
ow rate as follows:
From Main Menu:
Touch Sample
> Touch Flow
> Touch
calibration icon
Pump will run as a result of touching the calibration icon. It is good practice to allow the pump to
run for 5 minutes before calibrating the
fl
ow rate.
Build a calibration train with sample media in line
(Figure 3).
Touch up/down arrow bu
tt
ons to increment/decrement calibration adjustment.
Pump inlet
222 Series (non-adjustable) Low Flow Tube
Holder
Sorbent tube
Tubing
Calibrator outlet
Figure 3.
Single-tube Calibration Train Overview
Sorbent tube Tubing Calibrator ou
S
C