Universal Pump Service Manual
1
1
1
1
1
Introduction
Universal XR Sample Pumps
Universal XR Sample Pumps are pneumatic constant fl ow samplers that contain a fully electronic system built for
a broad range of applications. Flow is compensated to within ± 5% accuracy over the full range of fl ows and back
pressure.
There are three Universal XR pump models:
224-44XR
No additional features
224-PCXR4
LCD and 2-bu on keypad for built-in timer and fl ow fault/ba ery check
224-PCXR8
LCD and 6-bu on keypad for programming delayed start time, automatic shutdown
time, and intermi ent sampling plus fl ow fault/ba ery check
Models 44XR, PCXR4, and PCXR8 are RFI/EMI-shielded and UL Listed for intrinsic safety.
Performance Profi le
Operating Range:
1000 to 5000 ml/min (5 to 500 ml/min requires adjustable low fl ow holder)
Weight:
34 oz (964 gm)
Dimensions:
5.1 x 4.7 x 1.9 inches; (13 x 11.9 x 4.8 cm)
Compensation
Range:
1000 ml/min up to 40 inches water back pressure
2000 ml/min up to 40 inches water back pressure
2500 ml/min up to 40 inches water back pressure
3000 ml/min up to 35 inches water back pressure
4000 ml/min up to 20 inches water back pressure
5000 ml/min up to 10 inches water back pressure
Flow Control:
Holds constant fl ow to ± 5% of the set point
Run Time:
8 hrs minimum at 4000 ml/min and 20 in water back pressure, dependent on media used
Flow Indicator:
Built-in rotameter with 250 ml division; scale marked at 1, 2, 3, 4, and 5 L/min
Power Supply:
Charging Time:
(with SKC approved charger)
Plug-in ba ery pack, rechargeable NiCad 2.0-Ah capacity, 6.0 V, UL Listed
6 to 8.5 hrs with PowerFlex Charger
Operating Temp:
32 to 113 F (0 to 45 C)
Storage Temp:
-40 to 113 F (-40 to 45 C)
Charging Temp:
50 to 113 F (10 to 45 C)
Operating Humidity:
0 to 95% non-condensing
Multiple Sampling:
Built-in constant pressure regulator allows user to take up to four simultaneous samples
at diff erent fl ow rates up to 500 ml/min each using optional adjustable low fl ow holder.
RFI-shielding
Performance:
Complies with requirements of EN 55022, FCC Part 15 Class B, EN 50082-1; frequency
range of the radiated susceptibility test was 27 MHz to 1000 MHz, CE marked.
Intrinsic Safety:
UL Listed for: Class I, Division 1 and 2, Groups A, B, C, D; Class II, Division 1 and 2,
Groups E, F, G; and Class III, Temp Code T3C.