![SKC 224-44XR Скачать руководство пользователя страница 30](http://html.mh-extra.com/html/skc/224-44xr/224-44xr_service-manual_1286832030.webp)
26
26
26
26
26
Addendum B
224-44XRM, 224-PCXR4M, and 224-PCXR8M
(4 Cell) Pump Calibration
Model ____________
Serial Number _____________
Date: _________
Settings
Readings
Flow
BP”
Minimum
Flow
Maximum
Max
4000
0
4000
4000
4000
8
3800
4200
3000
0
3000
3000
3000
10
2850
3150
3000
15
2850
3150
3000
22
2850
3150
2500
0
2500
2500
2500
15
2375
2625
2500
25
2375
2625
2500
30
2375
2625
2000
0
2000
2000
2000
15
1900
2100
2000
30
1900
2100
2000
35
1900
2100
2000
40
1900
2100
1000
0
1000
1000
1000
15
950
1050
1000
30
950
1050
1000
35
950
1050
1000
40
950
1050
Check Points
ü
Maximum Flow
____
Set Regulator 20”
____
Keypad
____
Flow Fault
____