Universal Pump Service Manual
15
15
15
15
15
Troubleshooting
• Use a known good control board, stack, and motor when troubleshooting. SKC also recommends that a power supply be
used for all testing. Connect it directly to the ba ery connector on the control board a er removing the ba ery pack.
Use only SKC-approved parts to ensure reliable performance. Failure to do so voids UL Listing for
intrinsic safety and any warranty.
Failure to follow warnings and cautions voids any warranty.
Pump
Problem
Possible Solution
224-PCXR4 and PCXR8
Pump LCD doesn’t
light when pump is
turned on and the
motor will not run.
• Check that motor wiring is undamaged and solder joints at the
motor and on the board are in good repair.
• Check the battery for charge and proper voltage.
• Swap with a known good battery for further troubleshooting.
224-PCXR4 and PCXR8
Pump LCD lights
when pump is
turned on, but the
motor will not run.
Turn the fl ow adjust pot clockwise a few turns. The pot may have
been adjusted to the lowest setting.
• If the motor still does not respond, follow steps above.
• If the motor runs at full speed, turn the fl ow adjust pot
counterclockwise.
• If the pot cannot be adjusted clockwise, it may have separated
from the control board. Check control board. If pot has
separated, resolder pot and test operation (
see Functional
Testing, Flow Rate Spot Check
). If pot doesn't operate properly,
replace the control board (Cat. Nos. P22433U [PCXR8] or
P22433T [PCXR4]).
224-PCXR4 and PCXR8
Pump runs when
turned on, but the
LCD will not light.
Check the battery voltage or switch to an alternate power supply
to check the LCD. If it still doesn’t light, it is either a bad LCD
display or a bad diode on the control board. Replace LCD (Cat.
No. P72380,
see Pump Disassembly and Pump Reassembly
)
or replace the control board (Cat. Nos. P22433U [PCXR8] or
P22433T [PCXR4]).