61 30 459 D 3497
20
D 3497
.
201.01.04
.
02 06.2009
6 Operation
Sirona Dental Systems GmbH
Programs
Operating Instructions inFire HTC
Graphical representation
This example describes the programming of program number 11 with the
values of the table shown above.
1. Press the Load Program key.
2. Enter the program number
11
.
3. Press the S2 key (=
YES
) to confirm the loading procedure.
Programming step S3
1. Press the S3 key. The cursor will blink in the "Heating velocity" column of
line S3.
2. Enter the value
12
. The cursor will blink in the "Temperature" column.
3. Enter the value
1540
. The cursor will blink in the "Holding time" column.
4. Enter the value
120
.
Programming step S4
➢ Enter the values for step 4 as shown in the table.
Steps S1 and S2
Since steps S1 and S2 are not needed in this example, you can leave their
values at
0
with the exception of the heating velocity.
t
°C
1540
300
120
min
NOTICE
Entries with less than four digits
If you did not enter the temperature as
0300
and used
300
instead, the
cursor must be manually moved to the "Holding time" column by pressing
the S1 key.
Содержание inFire HTC