Basically, all connections are protected against external interference.
Choose a place of operation that excludes inductive or capacitive interfer-
ence influences on the magnetic sensor. When mounting the system keep a
maximum possible distance from lines loaded with interference. If neces-
sary, provide additional installations including screening shields or met-
allized housings. Contactor coils must be linked with spark suppression.
Connection
This magnetic sensor has been prepared for connection to SIKO electronic
display MA100/2 or MA523/1. No modification of the sensor connection
(e. g. cable modification) is permitted.
1. Bring the connector into position before the receptacle.
2. Pull backwards the locking
1
of the connector.
3. Plug the connector into the receptacle to the end stop.
4. Release the locking
1
.
1
Locking (black
plastic part)
1
Fig. 8: Connection
5 Commissioning
Following correct mounting and wiring you can put the measuring system
into operation by switching on operating voltage.
6 Trouble shooting
Below are some typical errors which may occur during installation and
operation:
•
Magnetic tape wrongly mounted, active side below (see chapter
).
•
The masking tape provided was not used for protecting the magnetic
tape. The masking tape must not be magnetizable.
•
Magnetic sensor not or incorrectly connected.
•
The distance tolerances were not observed over the whole measuring
length (see
).
•
Cable squeezed / interrupted / cut by sharp edges.
•
The magnetic sensor has been mounted with the active side turned
away from the band (see
NOTICE
NOTICE