UM RTD.2
Related Functions
Technical changes reserved
- 21 -
4.3 Calibrated sensors (point correction)
The calibrated sensors function of the device makes it possible to use sensors, the calibration (correc-
tion) coefficients of which are taken into consideration by the device during measurement.
•
Using the F1 key, enter the Configuration menu.
•
Select the
Calibrated sensors
function, followed by one of the 5 available sensors.
•
Confirm by pressing ENTER.
•
Enter the sensor information fields. Use the F1 function key (
) to move from one field to the
next.
•
Confirm your choice using the
VAL
key.
•
To add a value in the table of calibration points, use the
keys, enter the calibration points
(real value and value read) then confirm using the
VAL
key.