SIKA Dr.Siebert & Kühn GmbH & Co.KG
The line 5
difference
is always displayed. Depending on the equipment and selection, the temperature
difference is displayed as follows:
Without internal measuring instrument or with the internal measuring instrument switched off:
Difference = block temperature (internal reference) – temperature set-point (Fig. 6)
Fig. 6
Fig. 7
Fig. 8
With the internal measuring instrument switched on and
Probe selected and external reference connected:
Difference = probe – external reference (Fig. 5a)
Probe selected and external reference not connected:
Difference = probe – block temperature (internal reference) (Fig. 8)
Probe not selected and external reference connected:
Difference = external reference - temperature set-point (Fig. 7)
Probe not selected and external reference not connected:
Difference = block temperature (internal reference) - temperature set-point (Fig. 6)
The line
variance
is displayed only if selected in the “select display” submenu. It specifies by which value the
controlled block temperature varies.
Status Bar
The status bar can display four different statuses:
heating
when the temperature in the block is increased
cooling
when the block is cooled
stable
when the temperature in the block has set to a constant value
temp-alarm
when an alarm value/switch-off value set in the "Select System" submenu has
been
exceeded
When the internal measuring instrument is switched off and on again, all the probes are switched to "OFF".