4 Structure of the temperature calibrator
SIKA Dr.Siebert & Kühn GmbH & Co.KG
Software
Program package for TP 38… E
Remote control of the calibrator functions via PC
Setting of the display values (°C / °F / K, variance, min. and max. value, filter)
Programmable ramps, level and temperature cycle functions
System settings via PC (language, date, time, brightness, contrast, alarm)
Editing of the test data in graphical and ASCII code
Program package for TP 38…
Remote control of the calibrator functions via PC
Setting of the display values (°C / °F / K, variance, min. and max. value, filter)
Programmable ramps, level and temperature cycle functions
System settings via PC (language, date, time, brightness, contrast, alarm)
Editing of the test data in graphical and ASCII code
Processing of probe data
RTD
Pt100/Pt500/Pt1000
TC K/J/N/E/R/T/B/S/L/U
Current
0(4)...20
mA
Temperature
switch
programming and interpretation of step- and logger function
automatic creation of certificates incl. the integration of customer data
Serial test (e.g. for incoming goods inspection)
External calibration reference
Hand sensor appropriate for TP 38… with internal measuring instrument
Customer-specific linearization through EEPROM
Measuring range
-50...550 °C
Tolerances
+/-0.05 °C in the range of –30.00 °C...199.99 °C
+/-0.3 °C in the range of 200.00 °C...550.00 °C
Corrosion and acid resistant stainless steel 1.4571
Dimensions: L = 300 mm, D = 3 mm
Robust plastic hand grip (150 mm)
Electric connection via PVC cable (1 m) with 7-pin SIKA connector
Other configurations available on demand.
4.10 DKD Calibration Certificate
The DKD calibration certificate gives you the certainty that the calibration results are based on national
standards and the measuring deviations do not exceed the error limits specified.
DKD calibrations are optionally available for all temperature calibrators.
The calibration certificate of a DKD calibration laboratory confirms that the dry well calibrator fulfills the high
demands placed on the calibration capabilities of a calibrator of that kind. These requirements are defined in
directive DKD-R 5-4.
To check the calibration of dry well calibrators, we recommend periodical measurements using a calibrated
thermometer. Without verification measurements using a calibrated thermometer, we strongly recommend
annual recalibration of the dry well calibrator.