SIKA Dr.Siebert & Kühn GmbH & Co.KG
6.5.1 Select RTD menu
This menu item is displayed if an internal measuring instrument is available and switched on.
After selecting the "calibr RTD" menu item in Select mode using the cursor key and confirming your selection
with the Select key, the "Select RTD" submenu will be displayed.
The settings for the connected resistance thermometer are made in this menu.
Make your selection using the cursor keys.
Activate your selection using the Select key.
If the check or calibration of an RTD resistance thermometer is enabled with "ON", operation will be initiated in
"calibr RTD” mode.
Fig. 10: Submenu „select RTD“
Switching ON/OFF
In the second line you can switch the resistance thermometer check on or off. In the third line you can select
the type of probe.
Probe Types
PT100
Pt 100 resistance thermometer
in 2 or 3 or 4 wire connection
PT500
Pt 500 resistance thermometer
in 2 or 3 or 4 wire connection
PT1000
Pt 1000 resistance thermometer
in 2 or 3 or 4 wire connection
Connection
The "4 mm terminals"-connection does not include setting options since resistance thermometers can only be
connected to 4 mm terminals.