Decommissioning and disposal / return
SIGRIST-PHOTOMETER AG
27 / 176
9
Troubleshooting
9.1
Isolate faults
Detailed information on troubleshooting can be found in the
gives an overview of possible troubleshooting.
Malfunction
Measure
No display
u
Check the service voltage.
Error message in display
u
Analyse error message (
Warning/error/priority messages [
Measured value seems wrong
u
Make sure that the sample to be measured corresponds to the operating
conditions.
u
Adjust calibration.
u
Check the correct mounting of the device.
u
Ensure that the servicing work has been carried out in accordance with
the servicing schedule.
u
Perform sensor check.
9.2
Warning/error/priority messages
Detailed information about the messages can be found in the
. The messages are divided
into the following categories.
Colour
Category
Description
Image
Orange
WARNING
Warnings call attention to an unusual condition.
They have the following effects:
l
System remains in operation.
l
Measurement results must be evaluated with
caution.
l
The cause of the warning should be corrected
in a timely manner.
l
Once the cause has been corrected, the warn-
ing is automatically deleted.
1
2
3
4
(1)
Error code*
(2)
Message
(3)
Measured value
(4)
Device information
*Abbreviation:
W= Warning
E= Error
P= Prioritised error
Red
ERROR
Errors indicate a malfunction which makes correct
measured value acquisition impossible. They have
the following effects:
l
Measured values are set to 0.
l
Assigned power output terminals go to the
programmed measured current
«On error.»
l
Assigned limit values are deactivated.
l
If an output is programmed for errors, it is
switched.
Red
PRIO
Prioritised errors draw attention to a serious mal-
function. Further operation is impossible and the
measured values are set to 0.
Prioritised errors can only be cleared by a
service technician.
10
Decommissioning and disposal / return
Decommissioning and disposal
The photometer and the associated peripheral devices must be disposed of in accordance with the regional legal
regulations. A list of the materials used for this device can be found in the
Return to the appropriate country representative
For all devices and spare parts that are returned, a completed RMA form must be sent to the responsible SIGRIST-
PHOTOMETER AG country representative (RMA form 14711D can be downloaded from
).
DE
EN
FR
ES
NL
PT
ZH
RU
CS
JA
IT
PL
Содержание TurBiScat 2 Ex
Страница 88: ...88 176 SIGRIST PHOTOMETER AG l l l l l l l l l l l l l 1 4 u 1 5 1 6...
Страница 94: ...94 176 SIGRIST PHOTOMETER AG 4 3 0 2 mm2 100 m 10 5 TOUCH WiFi 5 1 l 0 5s l 1 2s 5 cm 6 u u...
Страница 99: ...SIGRIST PHOTOMETER AG 99 176 u u u RMA u u u u u RMA RMA DE EN FR ES NL PT ZH RU CS JA IT PL...
Страница 102: ...102 176 SIGRIST PHOTOMETER AG 1 3 TurBiScat 2 Ex 1 Ex db IIC T3 T4 T5 T6 Ga Gb l l l l l l l l l l l l l 1 4 u 1 5 1 6...
Страница 112: ...112 176 SIGRIST PHOTOMETER AG u u 5 u 4 3 u 3 u 1 u 6 u 7 u u 2 1 3 4 5 6 2 7...
Страница 114: ...114 176 SIGRIST PHOTOMETER AG 10 101 RMA SIGRIST PHOTOMETER AG RMA 14711D www photometer com u u u RMA u u u u u RMA RMA...
Страница 132: ...132 176 SIGRIST PHOTOMETER AG Ex u u 24 VDC 100 240 VAC u 133 u u u u 130 u u u u u u u DIN EN 61010 1 u u u...
Страница 140: ...140 176 SIGRIST PHOTOMETER AG VARINLINE 10 renata batteries CR1025 8 2 u u u 5 u 4 3 u 3 u 1 u 6 u 7 u u 2 1 3 4 5 6 2 7...
Страница 142: ...142 176 SIGRIST PHOTOMETER AG u u u RMA u u u u u RMA RMA...
Страница 173: ...SIGRIST PHOTOMETER AG 173 176...
Страница 174: ...174 176 SIGRIST PHOTOMETER AG...
Страница 175: ...SIGRIST PHOTOMETER AG 175 176...