8
Signet 8900 Multi-Parameter
4.3 Output Module, optional; install zero (0) or one (1)
Mfr. Part No.
Description
3-8900.405-1
Two passive 4 to 20 mA outputs
3-8900.405-2
Two active 4 to 20 mA outputs
3-8900.405-3
Two 0 to 5/10 VDC outputs
•
Optional analog Outputs 1 & 2 are obtained via the I/O Module, but are not
required to be installed in order to use the optional analog output modules
listed immediately above.
•
The analog outputs obtained via these separate modules will always be
identifi ed as Outputs 3 & 4 in the 8900 menus. The slot on the rear panel is
labeled accordingly.
•
It is perfectly acceptable to mix and match analog output types between those
contained on the I/O Module and those obtained via these separate modules.
•
Any and all analog outputs are freely assignable to any channel.
•
All analog outputs available from the 8900 are isolated.
•
Passive outputs throttle current in loops powered by a supply external to the
8900.
•
Active outputs throttle current in loops powered by the 8900; no external
power source is required.
•
Voltage outputs are independently software selectable for operation from 0 to 5 or 0 to 10 VDC.
4.4 Relay Module, optional; install zero (0), one (1) or two (2)
Mfr. Part No.
Description
3-8900.403-1
Two dry-contact relays
3-8900.403-2
Two solid-state relays
3-8050.396
RC Filter Kit
•
Up to two sets of two relays may be installed directly into the 8900. The modules are
completely independent of one another, so it is not necessary for Relays 1 & 2 to be
installed in order to use Relays 3 & 4.
•
However, in accordance with the labeling on the rear panel, relays installed directly
into the 8900 via these plug-in modules will always be identifi ed in the 8900 menus as
Relays 1 & 2 and 3 & 4, respectively.
•
The red LED relay status indicators on the front panel of the 8900 are hardwired to
Relays 1 & 2. If the slot for Relays 1 & 2 is vacant, the red LEDs will be inoperative.
Status of all relays is available at all times in a single screen in View mode.
•
It is perfectly acceptable to mix and match the relay types installed directly in the 8900,
as well as between internal and external modules.
•
Any and all relays are freely assignable to any channel. Hysteresis
and time delay are adjustable for each relay.
•
Dry-contact relays are electromechanical switches with a moving
contact armature. They are suitable for many general-purpose
applications, AC or DC, including AC loads up to 250 V.
•
Solid-state relays are electronic switches with no moving parts. They
may be used with AC or DC loads, but have lower current handling
capability and voltage ratings than Dry-contact relays. Solid-state
relays will outlast Dry-contact relays in pulsing applications.
•
Switching active loads (most commonly inductive) can cause contact
arcing suffi cient to damage both types of relays. The RC Filter Kit or
"snubber" is available as an accessory to reduce or eliminate these
damaging effects.
3-8900.620
100−240VAC
50−60Hz, 24VA MAX
~
12−24 VDC
0.7 A MAX.
N
L
+
-
Power Connection
DO NOT attempt to connect both
AC and DC at the same time
C
NO
NC
NO
NC
C
Solid State Relays
Rating:
50 mA 30V
Mechanical Relays
Rating:
5A 250 VAC
5A 30 VDC
~
+
-
+
-
~
CAUTION
Содержание 8900
Страница 50: ...50 Signet 8900 Multi Parameter Notes ...