25
Signet 8900 Multi-Parameter
9.2 Channel Settings: If Channel Type = pH
T1 Units:
C >
Select the unit of measure for the temperature sensor inside the pH electrode:
°C
or
°F
pH1 Average:
Med >
Dampens display, output and relay response rates for this channel.
Off
: Near instantaneous updates;
Low
: 4 second averaging;
Med
: 8 second averaging;
Hi
: 32 second averaging
Select decimal resolution for pH TEMPERATURE display:
XXXXX. XXXX.X
XXX.XX
If this channel type is pH and the sensor is 4 to 20 mA, set the minimum and maximum
setpoints.
ORP1 Average:
Med >
9.3 Channel Settings: If Channel Type = ORP
If this channel type is ORP and the sensor is 4 to 20 mA, set the minimum and maximum
setpoints.
Dampens display, output and relay response rates for this channel.
Off
: Near instantaneous updates
Low
: 4 s
M
ed
: 8 s
Hi
: 32 s
T1 Decimal:
****.* >
4mA Set Point 1:
0.0000 pH >
20mA Set Point 1
14.000 pH >
4mA Set Point 1:
-1000 mV >
20mA Set Point 1
1000 mV >