49
Signet 8900 Multi-Parameter
Channel 1 ID:
Not Assigned?
Input Module
Not Found
Input/Output 1&2
Module Defective
Comm/Out 3&4
Module Defective
Cal Point Too
Close to Zero
Standard Too
Close to Slope
6 Sensors and
1 Ext Relay
Loading...
Please Wait
Found New
S3L Devices
Value Must Be
3240 Or Less
Value Must Be
0.1 Or Higher
Value Must Be
1 Or Higher
Value Must Be
400 Or Less
Value Must Be
0.05 Or Higher
Value Must Be
1000.0 Or Less
Value Must Be
14.00 Or Less
All Settings
Will Change!
Are You Sure?
Press <ENTER>
Requesting Data
From Host
Waiting For
Request...
Error:Invalid
Host Version
Error:No Data
From Host
Error: Data
Transfer Failed
Transmitting
Data...
Transfer
Complete
8900
Restarting
These messages indicate
component failure on the Comm
or I/O module.
A sensor is connected but has
not been assigned to a channel
Self explanatory.
Pressure calibration cannot be
done at zero.
pH calibration buffers must be
at least 2 pH units apart.
Messages displayed when
"LOAD S
3
L DEVICES" is
selected.
If a relay is in PWM mode, the
period must be greater tha 0.1
seconds, and less than 3240
seconds.
If a relay is in Proportional Pulse
mode, the mimimum pulse rate
is 1, the maximum rate is 400.
Relay pulse width is 0.05 ms minimum,
1000 ms maximum
In pH calibration, the buffers cannot exceed
14 pH in value.
When cloning the setup from a host unit
into a slave unit, these messages warn
of impending actions and verify that the
change is intentional.
If the host and slave units have incompatible
software versions.
Self explanatory.
Self explanatory.
Self explanatory.
Self explanatory.
Host unit is not transmitting to the slave.
Host unit is waiting for the slave to request
data download.
Slave unit is transmitting request for data
download from the host unit.