8860 Dual Channel Conductivity/Resistivity Controller 

"- - - -" 

"Value Must be 100 or less"

"Value must be more than 0"

"Value must be 400 or less"

"Reset to Factory Calibration"

"Too Much Error
CHECK SENSOR"

Display is over range.  This may be a 
normal condition if your process operates 
at/near the limits of the sensor range. 

The menu item being set is a percentage 
value and must be less than 100.

The menu item being set cannot be zero 
or a negative value. 

The Pulse Rate for Relay and Open 
Collector outputs cannot be greater than 
400 pulses per minute. 

A value of "0" is being set into
".. Set: Conductivity" menu item.

The calibration offset entered is 
beyond the allowable tolerances of the 
instrument. 

Check sensor for correct range.
Check Decimal setting in OPTIONS menu.
Check Calibrate menu settings for incompatible 
SOURCE and RANGE values.

Select a value from 0 to 100.
 

Select a value greater than zero. 

Select a pulse rate less than 400

This will remove any user calibration from the 
"..Set Conductivity" and "..Set Temperature" 
items in the Calibrate menu.

Check calibration procedure for accuracy.
Check sensor for proper operation.
Check any cable extensions for poor splices or 
termination.

Display Condition

Suggested Solutions

Possible Causes

Troubleshooting

Technical Note

If a Current Loop is locked at 3.6 mA, the problem is related to the temperature circuit:
This occurs only if the 8860 detects a resistance from the temperature sensor that is less than 250

Ω

 or greater than 2800

Ω

.

•  Check the sensor wiring for open/short or poor connections on white (TEMP IN) and black (ISO GND)wires.
•  The PT1000 temperature device in the sensor is defective.
•  The transmitter is defective.

Содержание 8860

Страница 1: ...00 to 240 VAC 10 50 60 Hz 20VA or 11 to 24 VDC 10 regulated 0 5 A max 3 8860 11 to 24 VDC 10 regulated 0 5 A max Three 4 to 20 mA Outputs Passive isolated fully adjustable and reversible 4 to 20 mA outputs are independently source selectable Max loop impedance 150 Ω 12 V 450 Ω 18 V 750 Ω 24 V Update rate 100 mS Accuracy 0 03 mA 25 C 24 VDC Open collector outputs 2 available optically isolated 50 m...

Страница 2: ...not insert two wires into a single terminal If necessary splice the wires together before inserting into the terminal RELAY 1 RELAY 2 RELAY 3 RELAY 4 N C N O COM N C N O COM N C N O COM N C N O COM RELAY 3 4 OPEN COLL 3 4 OPEN COLL 4 OPEN COLL 4 OPEN COLL 3 OPEN COLL 3 LOOP 2 LOOP 2 LOOP 3 LOOP 3 LOOP 1 LOOP 1 SHLD ISO GND TEMP 1 SGNL 1 SHLD ISO GND TEMP 2 SGNL 2 POWER OUTPUT OPTION L N 11 24V 0 5...

Страница 3: ...e process condition The maximum pulse rate is 400 pulses per minute Example As the process falls below 10 μS the output will start pulsing in relation to the process value the max pulse endpoint and the programmed pulses min Pulse rate will increase as the process approaches the programmed endpoint SHLD ISO GND TEMP 2 SGNL 2 SHLD ISO GND TEMP 1 SGNL 1 RED WHITE BLACK SILVER SHLD RED WHITE BLACK SI...

Страница 4: ...n Signet Conductivity Resistivity Transmitter C1 23 45µS C2 8 72µS Relay 1 Relay 2 ENTER VIEW menu During normal operation the ProcessPro displays the VIEW menu When using the CALIBRATE or OPTIONS menus the ProcessPro will return to the VIEW menu if no activity occurs for 10 minutes To select the item you want displayed press the UP or DOWN arrow keys The items will scroll in a continuous loop Cha...

Страница 5: ... UP DOWN keys in sequence After entering the Key Code the display will show the first item in the selected menu Step 3 Scroll menu with UP or DOWN arrow keys Step 4 Press RIGHT ARROW key to select menu item to be edited The first display element will begin flashing Step 5 Press UP or DOWN keys to edit the flashing element RIGHT ARROW key advances the flashing element Step 6 Press ENTER key to save...

Страница 6: ...onitored by Relay 1 Cond 1 Cond 2 Temp 1 Temp 2 Function Set Relay 1 activation point The maximum value acceptable is 999999 USP setpoints are high alarms where the setpoint is a percentage below the USP limit Relay 1 will be deactivated at setpoint Hysteresis depending on High or Low selection When the relay Mode is USP defined as a HIGH alarm Hysteresis is displayed in μS Set up to 6400 seconds ...

Страница 7: ... based on the measured temperature Using the USP function In the 8860 USP setpoints are defined as a percentage below the USP limit so a USP alarm is always a HIGH alarm The 8860 can be set to warn you if the conductivity approaches within a set percentage of the USP limit The following settings and conditions are required for a USP relay function 1 In the CALIBRATE menu RELAY MODE must be set to ...

Страница 8: ... 8 sec averaging if your process experiences frequent or extreme fluctuations Set the decimal to the best resolution for your application The display will automatically scale up to this restriction Select or Adjust the minimum and maximum current output The display value represents the precise current output Adjustment limits 3 80 mA 4 00 mA 5 00 mA 19 00 mA 20 00 mA 21 00 mA Use this setting to m...

Страница 9: ...onductivity Siemens 0 000020 Siemens Conductivity Sensor cell e g 0 1 Cell 0 000001 Siemens Simulation resistance Ω 100 000 Ω or 1μS cm 1 μS 1 X 10 6 Siemens or 0 000001 Siemens Connect the conductivity resistor between the Sgnl 1 and Iso Gnd or Sgnl 2 and Iso Gnd terminals Set Cond Adjust the conductivity value based on the resistor value See Editing Procedure and Calibrate menu Verify the linear...

Страница 10: ...lean water less than 0 2 μS Temperature effects are not linear in this range so the temperature coefficient is not easily determined This selection is recommended for all Resistivity applications measuring from 5 MΩ to 18MΩ This selection conforms to ASTM standard D1125 and D5391 Linear This selection allows you to calculate a custom temperature compensation value for Conductivity measurements in ...

Страница 11: ...mal setting in OPTIONS menu Check Calibrate menu settings for incompatible SOURCE and RANGE values Select a value from 0 to 100 Select a value greater than zero Select a pulse rate less than 400 This will remove any user calibration from the Set Conductivity and Set Temperature items in the Calibrate menu Check calibration procedure for accuracy Check sensor for proper operation Check any cable ex...

Страница 12: ...n Certified 3 2819 T2 159 000 082 CR Sanitary 0 01 Titanium 2 in 3 2819 T2C 159 000 084 CR Sanitary 0 01 Titanium 2 in Certified 3 2820 S1 159 000 089 CR Sanitary 0 1 SS 1 to 1 1 2 in 3 2820 S1C 159 000 091 CR Sanitary 0 1 SS 1 to 1 1 2 in Certified 3 2820 S2 159 000 090 CR Sanitary 0 1 SS 2 in 3 2820 S2C 159 000 092 CR Sanitary 0 1 SS 2 in Certified 3 2820 T1 159 000 624 CR Sanitary 0 1 Titanium ...

Отзывы: