8860 Dual Channel Conductivity/Resistivity Controller 

Notes on Steps 5 and 6:

•  All output functions remain active during editing.
•  Only the 

fl

 ashing element can be edited.

•  RIGHT ARROW key advances the 

fl

 ashing element in a continuous loop.

•  Edited value is effective immediately after pressing ENTER key.
•  If no key is pressed for 10 minutes unit will restore the last saved value and return to step 3.
•  Step 6 (pressing ENTER key) always returns you to Step 3.
•  Repeat steps 3-6 until all editing is completed.

Notes on Step 2:

If no key is pressed for 5 minutes while display is showing "Enter 
Key Code", the display will return to the VIEW menu.

Notes on Steps 3 and 4:

•  Refer to pages 6 and 7 for complete listing of menu items and their use.
•  From the Step 3 display, pressing the UP and DOWN keys simultaneously will 
return the display to the VIEW menu.
•  If no key is pressed for 10 minutes, display will also return to the VIEW menu.

ProcessPro Editing Procedure:

Step 1.  

Press and hold ENTER key:

 

 

•  2 seconds to select the CALIBRATE menu.

 

 

•  5 seconds to select the OPTIONS menu.

Step 2. 

  The Key Code is UP-UP-UP-DOWN keys in sequence.

 

 

•  After entering the Key Code, the display will show the 

fi

 rst item in the selected menu.

Step 3.

  Scroll menu with UP or DOWN arrow keys.

Step 4.  

Press RIGHT ARROW key to select menu item to be edited.

  • 

 

The 

fi

 rst display element will begin 

fl

 ashing.

Step 5.

  Press UP or DOWN keys to edit the 

fl

 ashing element.

 

 

• RIGHT ARROW key advances the 

fl

 ashing element.

Step 6. 

 Press ENTER key to save the new setting and return to Step 3.

OPTIONS

CALIBRATE

VIEW

2s

5s

Press &
hold for 
access:

ENTER

Step 5

Step 6

Notes on Step 1:

•  The View Menu is normally displayed.
•  The CALIBRATE and OPTIONS menus require a KEY CODE.

Step 4

First item in CALI-
BRATE menu

Step 3

Step 3:  Finished Editing?

Press the UP and DOWN keys simultaneously after 
saving the last setting to return to normal operation.

Press the UP and DOWN keys simultaneously 
while any element is 

fl

 ashing.  This will recall the 

last saved value of the item being edited and 
return you to Step 3.

Step 5: Made an Error?

20.00 u

Relay1 

Setpnt:

 

 

2

0.

00 

uS 

 

Relay1 

Setpnt:

1

0.

00 

uS 

ENTER

Relay1 Setpnt:

1

0

.00 uS  

Relay 

Setpnt:

1

9

.

00 

uS 

 

Relay1 

Setpnt:

Saving

 

Relay1 

Setpnt:

 

 

19.

00 

uS 

 

 

 

>

20.00 us   >

Relay 1 Setpnt:

Standard  

>

Chan 1 Cell:

CALIBRATE:

 

----

Enter 

Key 

Code

CALIBRATE:

 

*---

Enter 

Key 

Code

CALIBRATE:

 

**--

Enter 

Key 

Code

CALIBRATE:

 

***-

Enter 

Key 

Code

Standard  

>

Chan 1 Cell:

Содержание 8860

Страница 1: ...00 to 240 VAC 10 50 60 Hz 20VA or 11 to 24 VDC 10 regulated 0 5 A max 3 8860 11 to 24 VDC 10 regulated 0 5 A max Three 4 to 20 mA Outputs Passive isolated fully adjustable and reversible 4 to 20 mA outputs are independently source selectable Max loop impedance 150 Ω 12 V 450 Ω 18 V 750 Ω 24 V Update rate 100 mS Accuracy 0 03 mA 25 C 24 VDC Open collector outputs 2 available optically isolated 50 m...

Страница 2: ...not insert two wires into a single terminal If necessary splice the wires together before inserting into the terminal RELAY 1 RELAY 2 RELAY 3 RELAY 4 N C N O COM N C N O COM N C N O COM N C N O COM RELAY 3 4 OPEN COLL 3 4 OPEN COLL 4 OPEN COLL 4 OPEN COLL 3 OPEN COLL 3 LOOP 2 LOOP 2 LOOP 3 LOOP 3 LOOP 1 LOOP 1 SHLD ISO GND TEMP 1 SGNL 1 SHLD ISO GND TEMP 2 SGNL 2 POWER OUTPUT OPTION L N 11 24V 0 5...

Страница 3: ...e process condition The maximum pulse rate is 400 pulses per minute Example As the process falls below 10 μS the output will start pulsing in relation to the process value the max pulse endpoint and the programmed pulses min Pulse rate will increase as the process approaches the programmed endpoint SHLD ISO GND TEMP 2 SGNL 2 SHLD ISO GND TEMP 1 SGNL 1 RED WHITE BLACK SILVER SHLD RED WHITE BLACK SI...

Страница 4: ...n Signet Conductivity Resistivity Transmitter C1 23 45µS C2 8 72µS Relay 1 Relay 2 ENTER VIEW menu During normal operation the ProcessPro displays the VIEW menu When using the CALIBRATE or OPTIONS menus the ProcessPro will return to the VIEW menu if no activity occurs for 10 minutes To select the item you want displayed press the UP or DOWN arrow keys The items will scroll in a continuous loop Cha...

Страница 5: ... UP DOWN keys in sequence After entering the Key Code the display will show the first item in the selected menu Step 3 Scroll menu with UP or DOWN arrow keys Step 4 Press RIGHT ARROW key to select menu item to be edited The first display element will begin flashing Step 5 Press UP or DOWN keys to edit the flashing element RIGHT ARROW key advances the flashing element Step 6 Press ENTER key to save...

Страница 6: ...onitored by Relay 1 Cond 1 Cond 2 Temp 1 Temp 2 Function Set Relay 1 activation point The maximum value acceptable is 999999 USP setpoints are high alarms where the setpoint is a percentage below the USP limit Relay 1 will be deactivated at setpoint Hysteresis depending on High or Low selection When the relay Mode is USP defined as a HIGH alarm Hysteresis is displayed in μS Set up to 6400 seconds ...

Страница 7: ... based on the measured temperature Using the USP function In the 8860 USP setpoints are defined as a percentage below the USP limit so a USP alarm is always a HIGH alarm The 8860 can be set to warn you if the conductivity approaches within a set percentage of the USP limit The following settings and conditions are required for a USP relay function 1 In the CALIBRATE menu RELAY MODE must be set to ...

Страница 8: ... 8 sec averaging if your process experiences frequent or extreme fluctuations Set the decimal to the best resolution for your application The display will automatically scale up to this restriction Select or Adjust the minimum and maximum current output The display value represents the precise current output Adjustment limits 3 80 mA 4 00 mA 5 00 mA 19 00 mA 20 00 mA 21 00 mA Use this setting to m...

Страница 9: ...onductivity Siemens 0 000020 Siemens Conductivity Sensor cell e g 0 1 Cell 0 000001 Siemens Simulation resistance Ω 100 000 Ω or 1μS cm 1 μS 1 X 10 6 Siemens or 0 000001 Siemens Connect the conductivity resistor between the Sgnl 1 and Iso Gnd or Sgnl 2 and Iso Gnd terminals Set Cond Adjust the conductivity value based on the resistor value See Editing Procedure and Calibrate menu Verify the linear...

Страница 10: ...lean water less than 0 2 μS Temperature effects are not linear in this range so the temperature coefficient is not easily determined This selection is recommended for all Resistivity applications measuring from 5 MΩ to 18MΩ This selection conforms to ASTM standard D1125 and D5391 Linear This selection allows you to calculate a custom temperature compensation value for Conductivity measurements in ...

Страница 11: ...mal setting in OPTIONS menu Check Calibrate menu settings for incompatible SOURCE and RANGE values Select a value from 0 to 100 Select a value greater than zero Select a pulse rate less than 400 This will remove any user calibration from the Set Conductivity and Set Temperature items in the Calibrate menu Check calibration procedure for accuracy Check sensor for proper operation Check any cable ex...

Страница 12: ...n Certified 3 2819 T2 159 000 082 CR Sanitary 0 01 Titanium 2 in 3 2819 T2C 159 000 084 CR Sanitary 0 01 Titanium 2 in Certified 3 2820 S1 159 000 089 CR Sanitary 0 1 SS 1 to 1 1 2 in 3 2820 S1C 159 000 091 CR Sanitary 0 1 SS 1 to 1 1 2 in Certified 3 2820 S2 159 000 090 CR Sanitary 0 1 SS 2 in 3 2820 S2C 159 000 092 CR Sanitary 0 1 SS 2 in Certified 3 2820 T1 159 000 624 CR Sanitary 0 1 Titanium ...

Отзывы: