Calibration
If the current device temperature is different from the calibration temperature, the gain
correction due to temperature difference is computed using
∆𝐺
𝑡𝑒𝑚𝑝
= 𝐶(𝑏1)
1
(
𝑇 − 𝑇
0
)
+ 𝐶(𝑏1)
2
(𝑇 − 𝑇
0
)
2
.
∆𝐺
𝑡𝑒𝑚𝑝
is the gain correction,
𝐶(𝑏1)
1
and
𝐶(𝑏1)
2
are the first and second order temperature
gain coefficients respectively for RF band 1, and
𝑇
and
𝑇
0
are the current temperature and
calibration temperature respectively. Adding this correction to the previously calculated gain will
compensate the value for temperature deviation.
𝐺 = 𝐺(𝑟𝑓)
𝑎𝑏𝑠
+ 𝐺(𝑖𝑓) −
𝐴
(
5𝑑𝐵, 𝑟𝑓
)
𝑟𝑓
𝑎𝑡𝑡𝑒𝑛
− 𝐴(2𝑑𝐵)
𝑖𝑓
𝑎𝑡𝑡𝑒𝑛
+
∆𝐺
𝑡𝑒𝑚𝑝