SC40401-K10-15002A
2nd
27
Copyright 2018, SIGMAKOKI Co.,Ltd.
1
~
4) STG_UT1
~
4
:
Select the units for display
Set the units used to display position coordinates of OPTION_unit (JS-300, JB-400).
[Select item]
PULSE
:
Number of pulses
MICRO
:
Micron m units
DEGREE
:
Degrees units
5
~
8) DIVIDE1
~
4
:
Divide setting for SG-55M
Set Divide for each axis. (For SG-55M).
*)
With regard to the Driver and Servo-pack of non-SG-55M, set the number of divisions in the
Driver and Servo-pack.
[Select item]
1/2/4/5/8/10/20/25/40/50/80/100/125/200/250
9
~
12) PLS_RATE1
~
4
:
Pulse Rate setting(1
~
4axes)
Set the travel distance per 1 pulse for each axis.
*) Otherwise, it is not able to position correctly. (Setting Unit
:
:
:
:
0.1nm ) *1)
[Setting range]
1
~
1000000 (0.1nm
~
100
µ
m) *1)
*1) If “STG_UT1~4” setting is “DEGREE”, please setting Unit: 0.000001
°
°
°
°
.
13
~
16) MOVE1
~
4
:
Travel direction setting 1~4axis
Set + travel direction for each axis.
[Select item]
POS
:
Positive rotation
NEG
:
Negative rotation
17
~
20) ORG1 SEL
~
ORG4 SEL
:
Origin reset method setting
Set Origin reset method for each axis.
[Select item]
OFF
:
Origin reset OFF
MINI
:
ORG1
CENTER
:
ORG2
ORGS
:
ORG3
NORM
:
ORG4
ZPM
:
ORG5
ZPP
:
ORG6
*) Regarding each method, please refer to “Origin Rest Method”.